Loading…
C55 CRITICAL CARE: CRITICAL CARE IN LOW AND MIDDLE INCOME COUNTRIES: A New Global And Comprehensive Model For Icu Ventilator Performance Evaluation
Technical performances were assessed under two FIO2 levels (100%, 50%), three respiratory mechanics combinations (Normal: compliance [C]=70 mL/cmH2O / resistance [R]=5 cmH2O/mL/s; Restrictive: C=30/R=10; Obstructive: C=20/R=120), four leaks levels (from 0 to 12.5 L/min), and three levels of inspirat...
Saved in:
Published in: | American journal of respiratory and critical care medicine 2017-01, Vol.195 |
---|---|
Main Authors: | , |
Format: | Article |
Language: | English |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Technical performances were assessed under two FIO2 levels (100%, 50%), three respiratory mechanics combinations (Normal: compliance [C]=70 mL/cmH2O / resistance [R]=5 cmH2O/mL/s; Restrictive: C=30/R=10; Obstructive: C=20/R=120), four leaks levels (from 0 to 12.5 L/min), and three levels of inspiratory effort (P0.1=2, 4, and 8 cmH2O), using an automated test lung. Ergonomics were evaluated by 20 ICU physicians using a global model involving physiological measurements (heart rate, respiratory rate, tidal volume variability and eye-tracking), psychocognitive scales (SUS and NASA-TLX), and objective tasks completion. |
---|---|
ISSN: | 1073-449X 1535-4970 |