Loading…
A neutron detector that is insensitive to attendant γ rays
A neutron detector is described, whose operating principle is based on detecting charge-transfer effects of secondary recoil protons produced in hydrogen-containing materials under exposure to primary neutrons. The physical mechanism of the operation of this detector is presented and its sensitivity...
Saved in:
Published in: | Instruments and experimental techniques (New York) 2017-11, Vol.60 (6), p.771-774 |
---|---|
Main Author: | |
Format: | Article |
Language: | English |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A neutron detector is described, whose operating principle is based on detecting charge-transfer effects of secondary recoil protons produced in hydrogen-containing materials under exposure to primary neutrons. The physical mechanism of the operation of this detector is presented and its sensitivity is determined as a function of the neutron energy. The design of a detector that is insensitive to attendant γ rays is proposed. The serviceability of the detector has been confirmed experimentally. The detector may find application to tests in mixed fields of γ rays and neutrons. |
---|---|
ISSN: | 0020-4412 1608-3180 |
DOI: | 10.1134/S0020441217060197 |