Loading…
Preparation of Calibration Gas Mixtures by the Chromato-Desorption Method for Increasing the Accuracy of Quantitative Determination of Biogenic Pentane in Expired Air
The possibility of increasing the accuracy of preparing n-pentane calibration gas mixtures by using injection-type chromato-desorption microsystems is investigated. The possibility of obtaining calibration gas mixtures in the range of 4–12 mg/m 3 (δ max = 16%) is shown. It is established that the ch...
Saved in:
Published in: | Measurement techniques 2017-11, Vol.60 (8), p.848-852 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The possibility of increasing the accuracy of preparing n-pentane calibration gas mixtures by using injection-type chromato-desorption microsystems is investigated. The possibility of obtaining calibration gas mixtures in the range of 4–12 mg/m
3
(δ
max
= 16%) is shown. It is established that the change in the composition of the mixture during storage for three months is not more than 7%. |
---|---|
ISSN: | 0543-1972 1573-8906 |
DOI: | 10.1007/s11018-017-1281-5 |