A comparison of the i‐gel™ with the LMA‐Unique™ in non‐paralysed anaesthetised adult patients
Summary This study assessed two disposable devices; the newly developed supraglottic airway device i‐gel™ and the LMA‐Unique™ in routine clinical practice. Eighty patients (ASA 1–3) undergoing minor routine gynaecologic surgery were randomly allocated to have an i‐gel (n = 40) or LMA‐Unique (n = 40)...
Saved in:
Published in: | Anaesthesia 2009-10, Vol.64 (10), p.1118-1124 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Summary
This study assessed two disposable devices; the newly developed supraglottic airway device i‐gel™ and the LMA‐Unique™ in routine clinical practice. Eighty patients (ASA 1–3) undergoing minor routine gynaecologic surgery were randomly allocated to have an i‐gel (n = 40) or LMA‐Unique (n = 40) inserted. Oxygen saturation, end‐tidal carbon dioxide, tidal volume and peak airway pressure were recorded, as well as time of insertion, airway leak pressure, postoperative sore‐throat, dysphonia and dysphagia for each device. Time of insertion was comparable with the i‐gel and LMA‐Unique. There was no failure in the i‐gel group and one failure in the LMA‐Unique group. Ventilation and oxygenation were similar between devices. Mean airway pressure was comparable with both devices, whereas airway leak pressure was significantly higher (p |
---|---|