Accuracy of vivo carotid b-mode ultrasound conpared with pathological analysis: Intima-media thickening, lumen diameter, and cross-sectional area
BACKGROUND AND PURPOSE: This study aimed to determine the correlation of in vivo ultrasound measurements of intima-media thickening (IMT), lumen diameter, and cross-sectional area of the common carotid artery (CCA) with corresponding measurements obtained by gross pathology and histology. METHODS: S...
Saved in:
Published in: | Stroke (1970) 2001-07, Vol.32 (7), p.1520 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | BACKGROUND AND PURPOSE: This study aimed to determine the correlation of in vivo ultrasound measurements of intima-media thickening (IMT), lumen diameter, and cross-sectional area of the common carotid artery (CCA) with corresponding measurements obtained by gross pathology and histology. METHODS: Sixty-six moribund neurological patients (mean age 71 years) underwent B-mode ultrasound of the CCA a few days before death. During autopsy, carotid specimens were removed in toto. Carotid arteries were ligated and cannulated for injection of a hydrophilic embedding material under standardized conditions. The carotid bifurcation was frozen and cut manually in 3-mm cross slices. Digital image analysis was carried out to determine the diameter and the cross-sectional area of the frozen slices of the CCA. IMT was assessed by light microscope. Ultrasonic and planimetric data were compared. RESULTS: Mean measurements of lumen diameter and cross-sectional area were 7.13+/-1.27 mm and 0.496+/-0.167 cm(2), respectively, by ultrasound, and 7.81+/-1.45 mm and 0.516+/-0.194 cm(2), respectively, by planimetric analysis of the unfixed redistended carotid arteries (R(2)=0.389 and 0.497). The mean IMT was 1.005+/-0.267 mm by ultrasound and 0.67+/-0.141 mm histologically, resulting in a mean difference of -31%. CONCLUSIONS: Transcutaneous B-mode ultrasound provides a reliable approach for in vivo measurements of the cross-sectional area and, less exactly, of the lumen diameter of the CCA. Compared with histological results, in vivo ultrasound measurements of the IMT are systematically larger. |
---|---|
ISSN: | 0039-2499 1524-4628 |