Development of a position-sensitive CZT detector with coplanar grid electrode
The finite–element analysis was done to optimize position-sensing coplanar grid electrode geometry. CdZnTe crystals manufactured by PAM-XIAMEN with customized electrode geometry have been tested. After digitizing pre-amplifier signals, trapezoidal digital filter was adopted to enhance the signal to...
Saved in:
Published in: | Journal of radioanalytical and nuclear chemistry 2018-06, Vol.316 (3), p.1221-1225 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The finite–element analysis was done to optimize position-sensing coplanar grid electrode geometry. CdZnTe crystals manufactured by PAM-XIAMEN with customized electrode geometry have been tested. After digitizing pre-amplifier signals, trapezoidal digital filter was adopted to enhance the signal to noise ratio. Energy resolution of 2.7% FWHM at 662 keV γ-ray energy was obtained. The measured results show the dramatic improvement with position-sensing technique. |
---|---|
ISSN: | 0236-5731 1588-2780 |
DOI: | 10.1007/s10967-018-5890-5 |