Calorimetric Power Measurements in the EAST ECRH System
In this paper, the calorimetric power measurement method for electron cyclotron resonance heating system on EAST are presented. This method requires measurement of the water flow through the cooling circuits and the input and output water temperatures in each cooling circuit. Usually, the inlet wate...
Saved in:
Published in: | arXiv.org 2017-03 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | In this paper, the calorimetric power measurement method for electron cyclotron resonance heating system on EAST are presented. This method requires measurement of the water flow through the cooling circuits and the input and output water temperatures in each cooling circuit. Usually, the inlet water temperature is controlled to be stable to get more accurate results. The influence of the inlet water temperature change on the measurement results is analyzed for the first time in this paper. A novel temperature calibration method is proposed also. This kind of calibration method is accurate and effective, and can be easily implemented. |
---|---|