Comparison of two full field digital mammography systems: Image quality and radiation dose

Full field digital mammography (FFDM) has been progressively introduced in medical centers in recent years. The goal of this study was to compare the average glandular doses (AGD) and image quality of two FFDM systems namely IMS Giotto and General Electric Senographe Essential. The AGD was calculate...

Full description

Saved in:
Bibliographic Details
Main Authors: Merad, A., Saadi, S., Khelassi-Toutaoui, N.
Format: Conference Proceeding
Language:English
Subjects:
Citations: Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Full field digital mammography (FFDM) has been progressively introduced in medical centers in recent years. The goal of this study was to compare the average glandular doses (AGD) and image quality of two FFDM systems namely IMS Giotto and General Electric Senographe Essential. The AGD was calculated according to Dance et al by using the measured entrance surface Kerma, acquisition parameters and c,g,s factors. The image quality was assessed quantitatively by measuring the contrast to noise ratio (CNR) value using a PMMA homogenous phantom of 2, 4, 6 ad 8 cm thicknesses and a small contrast object (foil of 0.2 mm Al) Following the European protocol. The obtained values were used to calculate the Figure of Merit (FOM). Measures analysis showed that there is a significant difference value of AGD and CNR for thicknesses greater than 4 cm between the two FFDM systems. A comparison of the CNR values in this study meets those recommended in the European protocol. GE Senographe Essential system contributed the lowest AGD value while IMS Giotto had the highest CNR and FOM value.
ISSN:0094-243X
1551-7616
DOI:10.1063/1.5048169