Calibrating an interferometric laser frequency stabilization to MHz precision

We report on a calibration procedure that enhances the precision of an interferometer based frequency stabilization by several orders of magnitude. For this purpose the frequency deviations of the stabilization are measured precisely by means of a frequency comb. This allows to implement several cal...

Full description

Saved in:
Bibliographic Details
Published in:arXiv.org 2012-08
Main Authors: Brachmann, Johannes F S, Kinder, Thomas, Dieckmann, Kai
Format: Article
Language:English
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:We report on a calibration procedure that enhances the precision of an interferometer based frequency stabilization by several orders of magnitude. For this purpose the frequency deviations of the stabilization are measured precisely by means of a frequency comb. This allows to implement several calibration steps that compensate different systematic errors. The resulting frequency deviation is shown to be less than \(5.7 \)MHz (rms \(1.6 \)MHz) in the whole wavelength interval \(750 - 795 \)nm. Wide tuning of a stabilized laser at this exceptional precision is demonstrated.
ISSN:2331-8422