Calibration of second-order correlation functions for non-stationary sources with a multi-start multi-stop time-to-digital converter

A novel high-throughput second-order-correlation measurement system is developed which records and makes use of all the arrival times of photons detected at both start and stop detectors. This system is suitable particularly for a light source having a high photon flux and a long coherence time sinc...

Full description

Saved in:
Bibliographic Details
Published in:arXiv.org 2005-03
Main Authors: Choi, Wonshik, Lee, Moonjoo, Ye-Ryoung, Lee, Park, Changsoon, Jai-Hyung, Lee, An, Kyungwon, Fang-Yen, C, Dasari, R R, Feld, M S
Format: Article
Language:English
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by
cites
container_end_page
container_issue
container_start_page
container_title arXiv.org
container_volume
creator Choi, Wonshik
Lee, Moonjoo
Ye-Ryoung, Lee
Park, Changsoon
Jai-Hyung, Lee
An, Kyungwon
Fang-Yen, C
Dasari, R R
Feld, M S
description A novel high-throughput second-order-correlation measurement system is developed which records and makes use of all the arrival times of photons detected at both start and stop detectors. This system is suitable particularly for a light source having a high photon flux and a long coherence time since it is more efficient than conventional methods by an amount equal to the product of the count rate and the correlation time of the light source. We have used this system in carefully investigating the dead time effects of detectors and photon counters on the second-order correlation function in the two-detector configuration. For a non-stationary light source, distortion of original signal was observed at high photon flux. A systematic way of calibrating the second-order correlation function has been devised by introducing a concept of an effective dead time of the entire measurement system.
doi_str_mv 10.48550/arxiv.0411212
format article
fullrecord <record><control><sourceid>proquest</sourceid><recordid>TN_cdi_proquest_journals_2090522362</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2090522362</sourcerecordid><originalsourceid>FETCH-proquest_journals_20905223623</originalsourceid><addsrcrecordid>eNqNjbFOAzEQRK1ISEQhLfVK1A72-hySOgLxAfSRufOBI8cb1nsBej48F0DUVPOkN5pR6tqaRbPy3twG_kjHhWmsRYsTNUXnrF41iJdqXuvOGIPLO_TeTdXXJuT0zEESFaAeamypdJq4iwwtMcf84_qhtGeo0BNDoaKrfJvAn1Bp4DZWeE_yCgH2Q5Z09ix_TAeQtI9aSHfpJUnI43w5RpbIV-qiD7nG-W_O1M3D_dPmUR-Y3oZYZbsbD8qotmjWxiO6Jbr_tU5SOVll</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2090522362</pqid></control><display><type>article</type><title>Calibration of second-order correlation functions for non-stationary sources with a multi-start multi-stop time-to-digital converter</title><source>Publicly Available Content Database (Proquest) (PQ_SDU_P3)</source><creator>Choi, Wonshik ; Lee, Moonjoo ; Ye-Ryoung, Lee ; Park, Changsoon ; Jai-Hyung, Lee ; An, Kyungwon ; Fang-Yen, C ; Dasari, R R ; Feld, M S</creator><creatorcontrib>Choi, Wonshik ; Lee, Moonjoo ; Ye-Ryoung, Lee ; Park, Changsoon ; Jai-Hyung, Lee ; An, Kyungwon ; Fang-Yen, C ; Dasari, R R ; Feld, M S</creatorcontrib><description>A novel high-throughput second-order-correlation measurement system is developed which records and makes use of all the arrival times of photons detected at both start and stop detectors. This system is suitable particularly for a light source having a high photon flux and a long coherence time since it is more efficient than conventional methods by an amount equal to the product of the count rate and the correlation time of the light source. We have used this system in carefully investigating the dead time effects of detectors and photon counters on the second-order correlation function in the two-detector configuration. For a non-stationary light source, distortion of original signal was observed at high photon flux. A systematic way of calibrating the second-order correlation function has been devised by introducing a concept of an effective dead time of the entire measurement system.</description><identifier>EISSN: 2331-8422</identifier><identifier>DOI: 10.48550/arxiv.0411212</identifier><language>eng</language><publisher>Ithaca: Cornell University Library, arXiv.org</publisher><subject>Converters ; Correlation analysis ; Detectors ; Photon counters ; Photons</subject><ispartof>arXiv.org, 2005-03</ispartof><rights>Notwithstanding the ProQuest Terms and conditions, you may use this content in accordance with the associated terms available at http://arxiv.org/abs/physics/0411212.</rights><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://www.proquest.com/docview/2090522362?pq-origsite=primo$$EHTML$$P50$$Gproquest$$Hfree_for_read</linktohtml><link.rule.ids>780,784,25753,27925,37012,44590</link.rule.ids></links><search><creatorcontrib>Choi, Wonshik</creatorcontrib><creatorcontrib>Lee, Moonjoo</creatorcontrib><creatorcontrib>Ye-Ryoung, Lee</creatorcontrib><creatorcontrib>Park, Changsoon</creatorcontrib><creatorcontrib>Jai-Hyung, Lee</creatorcontrib><creatorcontrib>An, Kyungwon</creatorcontrib><creatorcontrib>Fang-Yen, C</creatorcontrib><creatorcontrib>Dasari, R R</creatorcontrib><creatorcontrib>Feld, M S</creatorcontrib><title>Calibration of second-order correlation functions for non-stationary sources with a multi-start multi-stop time-to-digital converter</title><title>arXiv.org</title><description>A novel high-throughput second-order-correlation measurement system is developed which records and makes use of all the arrival times of photons detected at both start and stop detectors. This system is suitable particularly for a light source having a high photon flux and a long coherence time since it is more efficient than conventional methods by an amount equal to the product of the count rate and the correlation time of the light source. We have used this system in carefully investigating the dead time effects of detectors and photon counters on the second-order correlation function in the two-detector configuration. For a non-stationary light source, distortion of original signal was observed at high photon flux. A systematic way of calibrating the second-order correlation function has been devised by introducing a concept of an effective dead time of the entire measurement system.</description><subject>Converters</subject><subject>Correlation analysis</subject><subject>Detectors</subject><subject>Photon counters</subject><subject>Photons</subject><issn>2331-8422</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2005</creationdate><recordtype>article</recordtype><sourceid>PIMPY</sourceid><recordid>eNqNjbFOAzEQRK1ISEQhLfVK1A72-hySOgLxAfSRufOBI8cb1nsBej48F0DUVPOkN5pR6tqaRbPy3twG_kjHhWmsRYsTNUXnrF41iJdqXuvOGIPLO_TeTdXXJuT0zEESFaAeamypdJq4iwwtMcf84_qhtGeo0BNDoaKrfJvAn1Bp4DZWeE_yCgH2Q5Z09ix_TAeQtI9aSHfpJUnI43w5RpbIV-qiD7nG-W_O1M3D_dPmUR-Y3oZYZbsbD8qotmjWxiO6Jbr_tU5SOVll</recordid><startdate>20050315</startdate><enddate>20050315</enddate><creator>Choi, Wonshik</creator><creator>Lee, Moonjoo</creator><creator>Ye-Ryoung, Lee</creator><creator>Park, Changsoon</creator><creator>Jai-Hyung, Lee</creator><creator>An, Kyungwon</creator><creator>Fang-Yen, C</creator><creator>Dasari, R R</creator><creator>Feld, M S</creator><general>Cornell University Library, arXiv.org</general><scope>8FE</scope><scope>8FG</scope><scope>ABJCF</scope><scope>ABUWG</scope><scope>AFKRA</scope><scope>AZQEC</scope><scope>BENPR</scope><scope>BGLVJ</scope><scope>CCPQU</scope><scope>DWQXO</scope><scope>HCIFZ</scope><scope>L6V</scope><scope>M7S</scope><scope>PIMPY</scope><scope>PQEST</scope><scope>PQQKQ</scope><scope>PQUKI</scope><scope>PTHSS</scope></search><sort><creationdate>20050315</creationdate><title>Calibration of second-order correlation functions for non-stationary sources with a multi-start multi-stop time-to-digital converter</title><author>Choi, Wonshik ; Lee, Moonjoo ; Ye-Ryoung, Lee ; Park, Changsoon ; Jai-Hyung, Lee ; An, Kyungwon ; Fang-Yen, C ; Dasari, R R ; Feld, M S</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-proquest_journals_20905223623</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2005</creationdate><topic>Converters</topic><topic>Correlation analysis</topic><topic>Detectors</topic><topic>Photon counters</topic><topic>Photons</topic><toplevel>online_resources</toplevel><creatorcontrib>Choi, Wonshik</creatorcontrib><creatorcontrib>Lee, Moonjoo</creatorcontrib><creatorcontrib>Ye-Ryoung, Lee</creatorcontrib><creatorcontrib>Park, Changsoon</creatorcontrib><creatorcontrib>Jai-Hyung, Lee</creatorcontrib><creatorcontrib>An, Kyungwon</creatorcontrib><creatorcontrib>Fang-Yen, C</creatorcontrib><creatorcontrib>Dasari, R R</creatorcontrib><creatorcontrib>Feld, M S</creatorcontrib><collection>ProQuest SciTech Collection</collection><collection>ProQuest Technology Collection</collection><collection>Materials Science &amp; Engineering Collection</collection><collection>ProQuest Central (Alumni)</collection><collection>ProQuest Central</collection><collection>ProQuest Central Essentials</collection><collection>ProQuest Central</collection><collection>Technology Collection</collection><collection>ProQuest One Community College</collection><collection>ProQuest Central Korea</collection><collection>SciTech Premium Collection (Proquest) (PQ_SDU_P3)</collection><collection>ProQuest Engineering Collection</collection><collection>Engineering Database</collection><collection>Publicly Available Content Database (Proquest) (PQ_SDU_P3)</collection><collection>ProQuest One Academic Eastern Edition (DO NOT USE)</collection><collection>ProQuest One Academic</collection><collection>ProQuest One Academic UKI Edition</collection><collection>Engineering collection</collection></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Choi, Wonshik</au><au>Lee, Moonjoo</au><au>Ye-Ryoung, Lee</au><au>Park, Changsoon</au><au>Jai-Hyung, Lee</au><au>An, Kyungwon</au><au>Fang-Yen, C</au><au>Dasari, R R</au><au>Feld, M S</au><format>book</format><genre>document</genre><ristype>GEN</ristype><atitle>Calibration of second-order correlation functions for non-stationary sources with a multi-start multi-stop time-to-digital converter</atitle><jtitle>arXiv.org</jtitle><date>2005-03-15</date><risdate>2005</risdate><eissn>2331-8422</eissn><abstract>A novel high-throughput second-order-correlation measurement system is developed which records and makes use of all the arrival times of photons detected at both start and stop detectors. This system is suitable particularly for a light source having a high photon flux and a long coherence time since it is more efficient than conventional methods by an amount equal to the product of the count rate and the correlation time of the light source. We have used this system in carefully investigating the dead time effects of detectors and photon counters on the second-order correlation function in the two-detector configuration. For a non-stationary light source, distortion of original signal was observed at high photon flux. A systematic way of calibrating the second-order correlation function has been devised by introducing a concept of an effective dead time of the entire measurement system.</abstract><cop>Ithaca</cop><pub>Cornell University Library, arXiv.org</pub><doi>10.48550/arxiv.0411212</doi><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier EISSN: 2331-8422
ispartof arXiv.org, 2005-03
issn 2331-8422
language eng
recordid cdi_proquest_journals_2090522362
source Publicly Available Content Database (Proquest) (PQ_SDU_P3)
subjects Converters
Correlation analysis
Detectors
Photon counters
Photons
title Calibration of second-order correlation functions for non-stationary sources with a multi-start multi-stop time-to-digital converter
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-01T06%3A12%3A38IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest&rft_val_fmt=info:ofi/fmt:kev:mtx:book&rft.genre=document&rft.atitle=Calibration%20of%20second-order%20correlation%20functions%20for%20non-stationary%20sources%20with%20a%20multi-start%20multi-stop%20time-to-digital%20converter&rft.jtitle=arXiv.org&rft.au=Choi,%20Wonshik&rft.date=2005-03-15&rft.eissn=2331-8422&rft_id=info:doi/10.48550/arxiv.0411212&rft_dat=%3Cproquest%3E2090522362%3C/proquest%3E%3Cgrp_id%3Ecdi_FETCH-proquest_journals_20905223623%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=2090522362&rft_id=info:pmid/&rfr_iscdi=true