Loading…

Modeling of Microfabricated Microneedles for Minimally Invasive Drug Delivery, Sampling and Analysis

We compare simulation to analysis and experiments for flows in three microneedle geometries--straight, bent and filtered. The bent microneedle was found to have the highest fluid carrying capacity of 0.082 ml/sec at 138 kPa with a Reynolds number of 738. A microneedle with a built in microfilter had...

Full description

Saved in:
Bibliographic Details
Published in:Biomedical microdevices 2003-09, Vol.5 (3), p.245
Main Authors: Trebotich, D, Zahn, J D, Prabhakarpandian, B, Liepmann, D
Format: Article
Language:English
Citations: Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:We compare simulation to analysis and experiments for flows in three microneedle geometries--straight, bent and filtered. The bent microneedle was found to have the highest fluid carrying capacity of 0.082 ml/sec at 138 kPa with a Reynolds number of 738. A microneedle with a built in microfilter had a flow rate of 0.07 ml/sec. Although the throughput of these microneedles is low they compare favorably with other microneedle designs. Laminar flow models were found to accurately predict the flow behavior through the microneedles. All computational modeling was performed with the CFDRC CFD-ACE + suite of software tools.
ISSN:1387-2176
1572-8781
DOI:10.1023/A:1025716427229