Influencing factors of viscosity measurement by rotational method
Rotational rheometer was used to measure the shear viscosity of a liquid reference material of viscosity named GBW13604. The influences of heat balance time, bath temperature, environment temperature and sampling volume on results of measurements were investigated. Furthermore, a comparison of accur...
Saved in:
Published in: | Polymer testing 2018-09, Vol.70, p.144-150 |
---|---|
Main Authors: | , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
cited_by | cdi_FETCH-LOGICAL-c395t-fad10af134ddf54fa5f5472f8e4213c67c3a251a992ac8517ff67eaa433a46373 |
---|---|
cites | cdi_FETCH-LOGICAL-c395t-fad10af134ddf54fa5f5472f8e4213c67c3a251a992ac8517ff67eaa433a46373 |
container_end_page | 150 |
container_issue | |
container_start_page | 144 |
container_title | Polymer testing |
container_volume | 70 |
creator | Zhang, Yinglan Huang, Mengmeng Kan, Ying Liu, Lianying Dai, Xiaoping Zheng, Guiling Zhang, Zhengdong |
description | Rotational rheometer was used to measure the shear viscosity of a liquid reference material of viscosity named GBW13604. The influences of heat balance time, bath temperature, environment temperature and sampling volume on results of measurements were investigated. Furthermore, a comparison of accuracy and repeatability of shear viscosity results between two commonly used systems with cone–plate and coaxial cylinder geometry was performed. Rheological tests indicate that the coaxial cylinder method is applicable to tests requiring high accuracy, while the cone-plate method is suitable for measurements that demand low sampling volume and fast speed.
•Studied effects of heat balance time, bath temperature, environment temperature and sample volume on viscosity measurement.•Compared accuracy and repeatability of shear viscosity results between cone–plate and coaxial cylinder geometries.•The measurement error generated by the influencing factors was quantified, which provided guidance for related fields. |
doi_str_mv | 10.1016/j.polymertesting.2018.06.034 |
format | article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_journals_2121672738</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S0142941818308109</els_id><sourcerecordid>2121672738</sourcerecordid><originalsourceid>FETCH-LOGICAL-c395t-fad10af134ddf54fa5f5472f8e4213c67c3a251a992ac8517ff67eaa433a46373</originalsourceid><addsrcrecordid>eNqNkDFPwzAQhS0EEqXwHyLBmuCznTiRWKqKlkqVWGC2jGODoyQutlMp_x5XZWFjuRvuvad7H0IPgAvAUD12xcH186B91CHa8bMgGOoCVwWm7AItoOY0J5TVl2iBgZG8YVBfo5sQOoxxmRIWaLUbTT_pUSV7ZqSKzofMmexog3LBxjkbtAyT14MeY_YxZ95FGa0bZZ8u8cu1t-jKyD7ou9-9RO-b57f1S75_3e7Wq32uaFPG3MgWsDRAWduakhlZpsmJqTUjQFXFFZWkBNk0RKq6BG5MxbWUjFLJKsrpEt2fcw_efU-psOjc5NMfQRAgUHHCaZ1UT2eV8i4Er404eDtIPwvA4gRNdOIvNHGCJnAlErRk35ztOjU5Wu1FUDbR0a31WkXROvu_oB_I23_A</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2121672738</pqid></control><display><type>article</type><title>Influencing factors of viscosity measurement by rotational method</title><source>ScienceDirect Journals</source><creator>Zhang, Yinglan ; Huang, Mengmeng ; Kan, Ying ; Liu, Lianying ; Dai, Xiaoping ; Zheng, Guiling ; Zhang, Zhengdong</creator><creatorcontrib>Zhang, Yinglan ; Huang, Mengmeng ; Kan, Ying ; Liu, Lianying ; Dai, Xiaoping ; Zheng, Guiling ; Zhang, Zhengdong</creatorcontrib><description>Rotational rheometer was used to measure the shear viscosity of a liquid reference material of viscosity named GBW13604. The influences of heat balance time, bath temperature, environment temperature and sampling volume on results of measurements were investigated. Furthermore, a comparison of accuracy and repeatability of shear viscosity results between two commonly used systems with cone–plate and coaxial cylinder geometry was performed. Rheological tests indicate that the coaxial cylinder method is applicable to tests requiring high accuracy, while the cone-plate method is suitable for measurements that demand low sampling volume and fast speed.
•Studied effects of heat balance time, bath temperature, environment temperature and sample volume on viscosity measurement.•Compared accuracy and repeatability of shear viscosity results between cone–plate and coaxial cylinder geometries.•The measurement error generated by the influencing factors was quantified, which provided guidance for related fields.</description><identifier>ISSN: 0142-9418</identifier><identifier>EISSN: 1873-2348</identifier><identifier>DOI: 10.1016/j.polymertesting.2018.06.034</identifier><language>eng</language><publisher>Barking: Elsevier Ltd</publisher><subject>Accuracy ; Cylinders ; Geometries ; Geometry ; Heat balance ; Polymer liquid ; Polymers ; Rheological properties ; Rotational method ; Sampling ; Shear viscosity ; Test procedures ; Viscometers ; Viscosity ; Viscosity measurement</subject><ispartof>Polymer testing, 2018-09, Vol.70, p.144-150</ispartof><rights>2018 Elsevier Ltd</rights><rights>Copyright Elsevier BV Sep 2018</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c395t-fad10af134ddf54fa5f5472f8e4213c67c3a251a992ac8517ff67eaa433a46373</citedby><cites>FETCH-LOGICAL-c395t-fad10af134ddf54fa5f5472f8e4213c67c3a251a992ac8517ff67eaa433a46373</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27923,27924</link.rule.ids></links><search><creatorcontrib>Zhang, Yinglan</creatorcontrib><creatorcontrib>Huang, Mengmeng</creatorcontrib><creatorcontrib>Kan, Ying</creatorcontrib><creatorcontrib>Liu, Lianying</creatorcontrib><creatorcontrib>Dai, Xiaoping</creatorcontrib><creatorcontrib>Zheng, Guiling</creatorcontrib><creatorcontrib>Zhang, Zhengdong</creatorcontrib><title>Influencing factors of viscosity measurement by rotational method</title><title>Polymer testing</title><description>Rotational rheometer was used to measure the shear viscosity of a liquid reference material of viscosity named GBW13604. The influences of heat balance time, bath temperature, environment temperature and sampling volume on results of measurements were investigated. Furthermore, a comparison of accuracy and repeatability of shear viscosity results between two commonly used systems with cone–plate and coaxial cylinder geometry was performed. Rheological tests indicate that the coaxial cylinder method is applicable to tests requiring high accuracy, while the cone-plate method is suitable for measurements that demand low sampling volume and fast speed.
•Studied effects of heat balance time, bath temperature, environment temperature and sample volume on viscosity measurement.•Compared accuracy and repeatability of shear viscosity results between cone–plate and coaxial cylinder geometries.•The measurement error generated by the influencing factors was quantified, which provided guidance for related fields.</description><subject>Accuracy</subject><subject>Cylinders</subject><subject>Geometries</subject><subject>Geometry</subject><subject>Heat balance</subject><subject>Polymer liquid</subject><subject>Polymers</subject><subject>Rheological properties</subject><subject>Rotational method</subject><subject>Sampling</subject><subject>Shear viscosity</subject><subject>Test procedures</subject><subject>Viscometers</subject><subject>Viscosity</subject><subject>Viscosity measurement</subject><issn>0142-9418</issn><issn>1873-2348</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2018</creationdate><recordtype>article</recordtype><recordid>eNqNkDFPwzAQhS0EEqXwHyLBmuCznTiRWKqKlkqVWGC2jGODoyQutlMp_x5XZWFjuRvuvad7H0IPgAvAUD12xcH186B91CHa8bMgGOoCVwWm7AItoOY0J5TVl2iBgZG8YVBfo5sQOoxxmRIWaLUbTT_pUSV7ZqSKzofMmexog3LBxjkbtAyT14MeY_YxZ95FGa0bZZ8u8cu1t-jKyD7ou9-9RO-b57f1S75_3e7Wq32uaFPG3MgWsDRAWduakhlZpsmJqTUjQFXFFZWkBNk0RKq6BG5MxbWUjFLJKsrpEt2fcw_efU-psOjc5NMfQRAgUHHCaZ1UT2eV8i4Er404eDtIPwvA4gRNdOIvNHGCJnAlErRk35ztOjU5Wu1FUDbR0a31WkXROvu_oB_I23_A</recordid><startdate>20180901</startdate><enddate>20180901</enddate><creator>Zhang, Yinglan</creator><creator>Huang, Mengmeng</creator><creator>Kan, Ying</creator><creator>Liu, Lianying</creator><creator>Dai, Xiaoping</creator><creator>Zheng, Guiling</creator><creator>Zhang, Zhengdong</creator><general>Elsevier Ltd</general><general>Elsevier BV</general><scope>AAYXX</scope><scope>CITATION</scope><scope>7SR</scope><scope>8FD</scope><scope>JG9</scope></search><sort><creationdate>20180901</creationdate><title>Influencing factors of viscosity measurement by rotational method</title><author>Zhang, Yinglan ; Huang, Mengmeng ; Kan, Ying ; Liu, Lianying ; Dai, Xiaoping ; Zheng, Guiling ; Zhang, Zhengdong</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c395t-fad10af134ddf54fa5f5472f8e4213c67c3a251a992ac8517ff67eaa433a46373</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2018</creationdate><topic>Accuracy</topic><topic>Cylinders</topic><topic>Geometries</topic><topic>Geometry</topic><topic>Heat balance</topic><topic>Polymer liquid</topic><topic>Polymers</topic><topic>Rheological properties</topic><topic>Rotational method</topic><topic>Sampling</topic><topic>Shear viscosity</topic><topic>Test procedures</topic><topic>Viscometers</topic><topic>Viscosity</topic><topic>Viscosity measurement</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Zhang, Yinglan</creatorcontrib><creatorcontrib>Huang, Mengmeng</creatorcontrib><creatorcontrib>Kan, Ying</creatorcontrib><creatorcontrib>Liu, Lianying</creatorcontrib><creatorcontrib>Dai, Xiaoping</creatorcontrib><creatorcontrib>Zheng, Guiling</creatorcontrib><creatorcontrib>Zhang, Zhengdong</creatorcontrib><collection>CrossRef</collection><collection>Engineered Materials Abstracts</collection><collection>Technology Research Database</collection><collection>Materials Research Database</collection><jtitle>Polymer testing</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Zhang, Yinglan</au><au>Huang, Mengmeng</au><au>Kan, Ying</au><au>Liu, Lianying</au><au>Dai, Xiaoping</au><au>Zheng, Guiling</au><au>Zhang, Zhengdong</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Influencing factors of viscosity measurement by rotational method</atitle><jtitle>Polymer testing</jtitle><date>2018-09-01</date><risdate>2018</risdate><volume>70</volume><spage>144</spage><epage>150</epage><pages>144-150</pages><issn>0142-9418</issn><eissn>1873-2348</eissn><abstract>Rotational rheometer was used to measure the shear viscosity of a liquid reference material of viscosity named GBW13604. The influences of heat balance time, bath temperature, environment temperature and sampling volume on results of measurements were investigated. Furthermore, a comparison of accuracy and repeatability of shear viscosity results between two commonly used systems with cone–plate and coaxial cylinder geometry was performed. Rheological tests indicate that the coaxial cylinder method is applicable to tests requiring high accuracy, while the cone-plate method is suitable for measurements that demand low sampling volume and fast speed.
•Studied effects of heat balance time, bath temperature, environment temperature and sample volume on viscosity measurement.•Compared accuracy and repeatability of shear viscosity results between cone–plate and coaxial cylinder geometries.•The measurement error generated by the influencing factors was quantified, which provided guidance for related fields.</abstract><cop>Barking</cop><pub>Elsevier Ltd</pub><doi>10.1016/j.polymertesting.2018.06.034</doi><tpages>7</tpages></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0142-9418 |
ispartof | Polymer testing, 2018-09, Vol.70, p.144-150 |
issn | 0142-9418 1873-2348 |
language | eng |
recordid | cdi_proquest_journals_2121672738 |
source | ScienceDirect Journals |
subjects | Accuracy Cylinders Geometries Geometry Heat balance Polymer liquid Polymers Rheological properties Rotational method Sampling Shear viscosity Test procedures Viscometers Viscosity Viscosity measurement |
title | Influencing factors of viscosity measurement by rotational method |
url | http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-11T17%3A59%3A43IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Influencing%20factors%20of%20viscosity%20measurement%20by%20rotational%20method&rft.jtitle=Polymer%20testing&rft.au=Zhang,%20Yinglan&rft.date=2018-09-01&rft.volume=70&rft.spage=144&rft.epage=150&rft.pages=144-150&rft.issn=0142-9418&rft.eissn=1873-2348&rft_id=info:doi/10.1016/j.polymertesting.2018.06.034&rft_dat=%3Cproquest_cross%3E2121672738%3C/proquest_cross%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c395t-fad10af134ddf54fa5f5472f8e4213c67c3a251a992ac8517ff67eaa433a46373%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=2121672738&rft_id=info:pmid/&rfr_iscdi=true |