Separation and determination of uranium in phosphoric acid medium using high-performance ion chromatography

A new method is presented for measuring of uranium at low levels in crude phosphoric acid without preconcentration or complexing agents using Dionex DX-500 ion chromatograph. Two isocratic elutions namely; 7 mM PDCA (for eluting the interfering cations) and 0.3 M (NH 4 ) 2 SO 4 /0.1 M HCl (for urani...

Full description

Saved in:
Bibliographic Details
Published in:Journal of radioanalytical and nuclear chemistry 2019-08, Vol.321 (2), p.413-419
Main Authors: El Gammal, Entesar M., Ahmed, Sherien H.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A new method is presented for measuring of uranium at low levels in crude phosphoric acid without preconcentration or complexing agents using Dionex DX-500 ion chromatograph. Two isocratic elutions namely; 7 mM PDCA (for eluting the interfering cations) and 0.3 M (NH 4 ) 2 SO 4 /0.1 M HCl (for uranium separation) with flow rate of 0.9 and 0.7 ml/min respectively. The optimized developed method has been validated by its application for the analysis of uranium in a certified sample (GBW04115).The developed method has been successfully applied to the determination of uranium in the prepared synthetic orthophosphoric acid and the crude phosphoric acid.