Loading…

Development of amperometric enzyme biosensor based on carbon fibre electrode and immobilized glucose oxidase

A glucose amperometric biosensor has been developed, its characteristics being studied. The stipulation of obtaining active membrane based on glucose oxidase has been optimized. The changes in both background electrolyte concentration and buffer capacity have been shown to affect the biosensor respo...

Full description

Saved in:
Bibliographic Details
Published in:Biopolimery i kletka 2003, Vol.19 (1), p.76-80
Main Authors: Danyleyko, L. V., Schuvailo, O. N., Arkhypova, V. M., Soldatkin, A. P., Dzyadevych, S. V.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A glucose amperometric biosensor has been developed, its characteristics being studied. The stipulation of obtaining active membrane based on glucose oxidase has been optimized. The changes in both background electrolyte concentration and buffer capacity have been shown to affect the biosensor response insufficiently. The optimum pH is found to be 7.2. The biosensor dynamic range is shown to be of 0–8 mM, the detection limit of 0.03 mM. The biosensor developed is recommended for the application in food industry to control and optimize biotechnological processes.
ISSN:0233-7657
1993-6842
DOI:10.7124/bc.00063E