Comparison of four analytical techniques for determining uranium in different rock samples
Four different analytical techniques were compared for their performance in direct determination of uranium in seven rock samples. These methods include: ion chromatography, laser fluorimetry, UV–Vis spectrophotometry and gamma spectrometry. The samples under study were six ore samples beside an int...
Saved in:
Published in: | Journal of radioanalytical and nuclear chemistry 2019-12, Vol.322 (3), p.1975-1982 |
---|---|
Main Authors: | , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Four different analytical techniques were compared for their performance in direct determination of uranium in seven rock samples. These methods include: ion chromatography, laser fluorimetry, UV–Vis spectrophotometry and gamma spectrometry. The samples under study were six ore samples beside an international reference sample (GBWO4115). The results obtained from the comparison study reported a method accuracy of 98%, 90%, 65% and 86% for ion chromatography, laser fluorimetry, UV–Vis spectrophotometry and gamma spectrometry techniques respectively. Also, on the converse of the other three techniques, ion chromatography technique using DX-500 ion chromatograph achieves superior agreement with the certified sample with only 2% bias. |
---|---|
ISSN: | 0236-5731 1588-2780 |
DOI: | 10.1007/s10967-019-06911-0 |