Detection of low energy electrons by a total absorption Cherenkov spectrometer

Presented are results of calibration, with a quasi-monochromatic electron beam, of a total absorption Cherenkov spectrometer based on the 14.8 X0 lead glass TF-1. It was found that the energy resolution of the spectrometer was 89% to 10% in the energy range of the electron beam E = 6 to 285 MeV, res...

Full description

Saved in:
Bibliographic Details
Published in:arXiv.org 2019-11
Main Authors: Alekseev, V I, Baskov, V A, Dronov, V A, L'vov, A I, Koltsov, A V, Krechetov, Yu F, Malinovsky, E I, Polyansky, V V
Format: Article
Language:English
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Presented are results of calibration, with a quasi-monochromatic electron beam, of a total absorption Cherenkov spectrometer based on the 14.8 X0 lead glass TF-1. It was found that the energy resolution of the spectrometer was 89% to 10% in the energy range of the electron beam E = 6 to 285 MeV, respectively.