Loading…
Test beam measurement of the first prototype of the fast silicon pixel monolithic detector for the TT-PET project
The TT-PET collaboration is developing a PET scanner for small animals with 30ps time-of-flight resolution and sub-millimetre 3D detection granularity. The sensitive element of the scanner is a monolithic silicon pixel detector based on state-of-the-art SiGe BiCMOS technology. The first ASIC prototy...
Saved in:
Published in: | Journal of instrumentation 2018-04, Vol.13 (4), p.P04015-P04015 |
---|---|
Main Authors: | , , , , , , , , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The TT-PET collaboration is developing a PET scanner for small animals with 30ps time-of-flight resolution and sub-millimetre 3D detection granularity. The sensitive element of the scanner is a monolithic silicon pixel detector based on state-of-the-art SiGe BiCMOS technology. The first ASIC prototype for the TT-PET was produced and tested in the laboratory and with minimum ionizing particles. The electronics exhibit an equivalent noise charge below 600e−RMS and a pulse rise time of less than 2ns, in accordance with the simulations. The pixels with a capacitance of 0.8pF were measured to have a detection efficiency greater than 99% and, although in the absence of the post-processing, a time resolution of approximately 200ps. |
---|---|
ISSN: | 1748-0221 1748-0221 |
DOI: | 10.1088/1748-0221/13/04/P04015 |