Loading…
Headspace-Gas Chromatography Method for Determination of Formaldehyde Content in Wastewater
Through the use of headspace-gas chromatography, the sensitivity and detection limit of formaldehyde content in wastewater obtained by the standard curve method and the standard addition method are compared in order to obtain the best detection method. The results show that the calibration curve cor...
Saved in:
Published in: | IOP conference series. Earth and environmental science 2021-01, Vol.647 (1), p.12155 |
---|---|
Main Authors: | , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Through the use of headspace-gas chromatography, the sensitivity and detection limit of formaldehyde content in wastewater obtained by the standard curve method and the standard addition method are compared in order to obtain the best detection method. The results show that the calibration curve correlation of the standard insertion method for testing the formaldehyde content in wastewater is 0.9991, the detection limit of the method is 0.18 mg/L, and the lower limit of determination is 0.69 mg/L. Excellent standard curve method; this method is used for samples with low, medium and high concentrations. The range of standard recovery rate calculated is 85.1%~112.9%, the relative standard deviation in the laboratory is 2.7%~ 6.3%, which is also better than the standard curve method and meets the test requirements. This method is easy to operate, high sensitivity, and low detection limit for the determination of formaldehyde in wastewater, which meets the requirements of analysis and testing. |
---|---|
ISSN: | 1755-1307 1755-1315 |
DOI: | 10.1088/1755-1315/647/1/012155 |