Supplementary comparison APMP.RI(I)-S1 of standards for absorbed dose to water in 60Co gamma radiation at radiation processing dose levels

Four national standards for absorbed dose to water in 60Co gamma radiation at the dose levels used in radiation processing have been compared over the range 0.1 kGy to 50 kGy using alanine dosimeters of the Office of Atoms for Peace (OAP) as transfer dosimeters. The comparison was piloted by the OAP...

Full description

Saved in:
Bibliographic Details
Published in:Metrologia 2021-01, Vol.58 (1A)
Main Authors: Mungpayaban, H, Ninlaphruk, S, Chen, C L, Yuan, M C, Zhang, Y L, Yuandi, Y, Hassan, G M, Rabie, N, Kadni, T B, Butler, D
Format: Article
Language:English
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Four national standards for absorbed dose to water in 60Co gamma radiation at the dose levels used in radiation processing have been compared over the range 0.1 kGy to 50 kGy using alanine dosimeters of the Office of Atoms for Peace (OAP) as transfer dosimeters. The comparison was piloted by the OAP who also participated. Two laboratories from the original six were forced to withdraw due to equipment problems. The results at low doses (0.1 kGy) showed a much wider spread (up to 17%) than at the other doses, most likely as a result of random variations in the alanine readout at OAP at these dose levels. Results in the 0.1 kGy range were excluded from the analysis because the variation between the laboratories' doses is overshadowed by the variation in the readout of the alanine. Above 1 kGy the results indicated reasonable agreement between the laboratories, with the majority of results within 2 % of the reference value. All of these results were within two combined standard uncertainties of the reference value, with the exception of the INER at 50 kGy point which was within three combined standard uncertainties. Within the stated uncertainties, the results establish the equivalence of the laboratories at radiation processing levels, for the range in which they participated: NIS (3 - 10) kGy, NIM (3 - 30) kGy, INER (3 - 50) kGy, OAP (3 - 50) kGy.To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/.The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).
ISSN:0026-1394
1681-7575
DOI:10.1088/0026-1394/58/1A/06021