Loading…

Radiation situation around 160 keV X-ray source

This paper presents the results of simulations concerning the radiation situation in and around the laboratory at FEI STU. These simulations focus on the X-ray source, which will be useful in future R&D activities. The purchased X-ray source is the SMART EVO 160D with a maximum photon energy of...

Full description

Saved in:
Bibliographic Details
Main Authors: Kleštinec Matej, Vrban Branislav, Lüley Jakub, Čerba Štefan, Nečas Vladimír
Format: Conference Proceeding
Language:English
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:This paper presents the results of simulations concerning the radiation situation in and around the laboratory at FEI STU. These simulations focus on the X-ray source, which will be useful in future R&D activities. The purchased X-ray source is the SMART EVO 160D with a maximum photon energy of 160 keV and uses a tungsten target. The model for the simulation was prepared in the SCALE 6.2 system and radiation doses were calculated using the MAVRIC sequence and variation reduction methods. The assumed energy of emitted photons is based on a tungsten anode energy spectrum. A discussion of the model specifications and the employed simplifications is provided.
ISSN:0094-243X
1551-7616
DOI:10.1063/5.0234903