A traceable precision current source for currents between 100 aA and 10 pA
A new calibration setup for picoamperemeters is presented. Owing to the use of a novel voltage ramp generator with increased linearity, the useful measuring time is increased to about 35 min, which reduces the calibration uncertainty of picomaperemeters with long time constants.
Saved in:
Published in: | IEEE transactions on instrumentation and measurement 2003-04, Vol.52 (2), p.436-439 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A new calibration setup for picoamperemeters is presented. Owing to the use of a novel voltage ramp generator with increased linearity, the useful measuring time is increased to about 35 min, which reduces the calibration uncertainty of picomaperemeters with long time constants. |
---|---|