Loading…

Count-rate dependent event mispositioning and NEC in PET

Most current PET detector designs suffer from event mispositioning at high count rates, as scintillation light from nearby and nearly simultaneous gamma ray conversions becomes mixed. We have used the NEMA NU 2-2001 70 cm test phantom and a Na-22 point source to quantify this effect as a function of...

Full description

Saved in:
Bibliographic Details
Published in:IEEE transactions on nuclear science 2004-02, Vol.51 (1), p.41-45
Main Authors: Badawi, R.D., Domigan, P., Johnson, O., Kemp, B., Kudrolli, H., Rempel, T., Rohatgi, R., Romanov, L.V., Surti, S., Worstell, W.A., Zimmerman, R.E.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Most current PET detector designs suffer from event mispositioning at high count rates, as scintillation light from nearby and nearly simultaneous gamma ray conversions becomes mixed. We have used the NEMA NU 2-2001 70 cm test phantom and a Na-22 point source to quantify this effect as a function of activity on two block-detector tomographs (the Siemens/CTI HR+ and the General Electric Discovery LS), and two Anger-type PET tomographs (the Siemens ECAM DUET and the Philips CPET+). After accounting for event losses due to dead time, we find that the number of counts on LORs passing through a cylinder of diameter and height equal to the point-source full width at tenth-maximum measured at low rate surrounding the point source decreases by between 9% (HR+) and 35% (CPET+) at the activity giving rise to peak noise equivalent count (NEC) rate. Mispositioned events act to reduce signal-to-noise ratio, both by reducing apparent activity at the originating location and by increasing the signal background. We have reformulated the conventional expression for NEC rate to account for this phenomenon. The new formulation of NEC, which we call NEC/sup */, results in a lower peak value which in turn occurs at a lower activity concentration than for the conventional formulation.
ISSN:0018-9499
1558-1578
DOI:10.1109/TNS.2003.823045