Loading…
Tritium Measurement in High Gamma-Ray Radiation Fields by Using an Imaging Plate
An imaging plate (IP) was applied to measure tritium in high 60 Co gamma-ray radiation fields. The IP made of europium-doped BaFBr(I), a photostimulated luminescence (PSL) material, is a two-dimensional radiation sensor. The PSL response of the IP has a peak at 20-50 keV and steeply decreases toward...
Saved in:
Published in: | Fusion science and technology 2011-10, Vol.60 (3), p.944-947 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | An imaging plate (IP) was applied to measure tritium in high
60
Co gamma-ray radiation fields. The IP made of europium-doped BaFBr(I), a photostimulated luminescence (PSL) material, is a two-dimensional radiation sensor. The PSL response of the IP has a peak at 20-50 keV and steeply decreases towards higher energy, falling by one hundredth at around 1 MeV. By utilizing a large difference in the PSL response to photon energy between
60
Co (1.173 and 1.333 MeV) and tritium (maximum energy of 18.6 keV), the bremsstrahlung X-ray induced by tritium beta ray was detected in mixed radiation fields with tritium and
60
Co, varying
60
Co dose rate in the range 0.0013 to 9.22 μGy/min. It was found that the effect of
60
Co irradiation to PSL value, obtained by irradiated with tritium of 12.5 MBq, was negligible by dose rate of 4.38 μGy/min and there was only 7.0% difference of PSL value, obtained by irradiated with tritium of 100 MBq, between dose rate of 0.0013 and 9.22 μGy/min. The IP tritium measurement method can be a promising candidate to measure tritium in high gamma-ray radiation fields. |
---|---|
ISSN: | 1536-1055 1943-7641 |
DOI: | 10.13182/FST11-A12571 |