Transcutaneous bilirubin measurement: Comparison of Respironics BiliCheck and JM-103 in a normal newborn population

To compare the accuracy of BiliCheck™ (Respironics, Marietta, GA) and Konica-Minolta Air Shield JM-103 (Drager Medical Inc, Telford, PA) to evaluate total serum bilirubin (TSB). Prospective blinded study comparing two diagnostic devices in 630 neonates requiring TSB measurement. Linear regression an...

Full description

Saved in:
Bibliographic Details
Published in:Clinical biochemistry 2012-06, Vol.45 (9), p.659-662
Main Authors: Romagnoli, Costantino, Zecca, Enrico, Catenazzi, Piero, Barone, Giovanni, Zuppa, Antonio Alberto
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:To compare the accuracy of BiliCheck™ (Respironics, Marietta, GA) and Konica-Minolta Air Shield JM-103 (Drager Medical Inc, Telford, PA) to evaluate total serum bilirubin (TSB). Prospective blinded study comparing two diagnostic devices in 630 neonates requiring TSB measurement. Linear regression analysis showed a good correlation between BiliCheck™ and TSB (r=0.8212) as well as between JM-103 and TSB (r=0.8686). BiliCheck shows a tendency to underestimate TSB. The mean difference in TSB–TcB was −1.4mg/dL for BC (−4.7/+1.8mg/dL) and 0.3mg/dL for JM-103 (−2.6/+3.2mg/dL). ROC analysis for TSB≥12mg/dL showed area under the curve for BiliCheck™ significantly lower than those for JM-103 (p