Transcutaneous bilirubin measurement: Comparison of Respironics BiliCheck and JM-103 in a normal newborn population
To compare the accuracy of BiliCheck™ (Respironics, Marietta, GA) and Konica-Minolta Air Shield JM-103 (Drager Medical Inc, Telford, PA) to evaluate total serum bilirubin (TSB). Prospective blinded study comparing two diagnostic devices in 630 neonates requiring TSB measurement. Linear regression an...
Saved in:
Published in: | Clinical biochemistry 2012-06, Vol.45 (9), p.659-662 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | To compare the accuracy of BiliCheck™ (Respironics, Marietta, GA) and Konica-Minolta Air Shield JM-103 (Drager Medical Inc, Telford, PA) to evaluate total serum bilirubin (TSB).
Prospective blinded study comparing two diagnostic devices in 630 neonates requiring TSB measurement.
Linear regression analysis showed a good correlation between BiliCheck™ and TSB (r=0.8212) as well as between JM-103 and TSB (r=0.8686). BiliCheck shows a tendency to underestimate TSB. The mean difference in TSB–TcB was −1.4mg/dL for BC (−4.7/+1.8mg/dL) and 0.3mg/dL for JM-103 (−2.6/+3.2mg/dL). ROC analysis for TSB≥12mg/dL showed area under the curve for BiliCheck™ significantly lower than those for JM-103 (p |
---|---|