Loading…
Orbscan II and double-K method for IOL calculation after refractive surgery
Background Precise IOL calculation in post-refractive surgery patients is still a challenge for the cataract surgeon. The purpose of this study is to test whether adding Orbscan II values into the double-K method improves IOL calculation in this group of patients. Methods A prospective study with 43...
Saved in:
Published in: | Graefe's archive for clinical and experimental ophthalmology 2012-07, Vol.250 (7), p.1029-1034 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Background
Precise IOL calculation in post-refractive surgery patients is still a challenge for the cataract surgeon. The purpose of this study is to test whether adding Orbscan II values into the double-K method improves IOL calculation in this group of patients.
Methods
A prospective study with 43 eyes previously submitted to refractive surgery that underwent cataract extraction. IOL calculation was performed with double-K method. Post-K value was derived from Orbscan total-mean power map. The average corneal curvature of the general population (43.8D) was used as the pre-K value. Refraction results 30 days after surgery were compared with refraction that would be obtained if we used: (1) post-K values from keratometry, (2) post-K values from topography, and (3) pre-K values from Orbscan total-mean power. Anterior chamber depth measures obtained with the IOL Master and Orbscan II were compared.
Results
Mean postoperative spherical equivalent (SE) was −0.25 ± 1.10 D in eyes submitted to radial keratotomy , –1.04 ± 1.42 D in eyes previously submitted to myopic Lasik, and +0.05 ± 1.76 D in those submitted to hyperopic surgeries. Had we inputted post-K values derived from keratometer and from topography, we would have obtained significantly higher postoperative refractive errors in eyes previously submitted to myopic refractive surgery (
p
0.05). Anterior chamber depth measured with IOL Master or Orbscan were similar.
Conclusions
Orbscan measurements used as the post-K values into the double-K method provide a precise IOL calculation, especially in post myopic refractive surgery patients. |
---|---|
ISSN: | 0721-832X 1435-702X |
DOI: | 10.1007/s00417-012-1974-z |