Loading…

An ICD Lead with Failure of Outer Insulation Goes Undetected by Regular Measurements

We report a patient with ischemic cardiomyopathy status post implantable cardioverter‐defibrillator (ICD) implantation in June 2009. Outer insulation failure of her Riata ST ventricular ICD lead goes undetected by routine interrogation and provocative tests as recommended from its manufacturer.1 Ano...

Full description

Saved in:
Bibliographic Details
Published in:Pacing and clinical electrophysiology 2012-09, Vol.35 (9), p.e261-e262
Main Authors: CHAN, CHI‐WO, CHIANG, CHUNG‐SEUNG
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:We report a patient with ischemic cardiomyopathy status post implantable cardioverter‐defibrillator (ICD) implantation in June 2009. Outer insulation failure of her Riata ST ventricular ICD lead goes undetected by routine interrogation and provocative tests as recommended from its manufacturer.1 Another domain of assessment, fluoroscopy, to facilitate early detection of structural abnormality in this family of ICD lead is discussed. (PACE 2012; 35:e261–e262)
ISSN:0147-8389
1540-8159
DOI:10.1111/j.1540-8159.2011.03164.x