Loading…

Image quality of ultra-low radiation exposure coronary CT angiography with an effective dose <0.1 mSv using high-pitch spiral acquisition and raw data-based iterative reconstruction

Objectives We evaluated the potential of prospectively ECG-triggered high-pitch spiral acquisition with low tube voltage and current in combination with iterative reconstruction to achieve coronary CT angiography with sufficient image quality at an effective dose below 0.1 mSv. Methods Contrast-enha...

Full description

Saved in:
Bibliographic Details
Published in:European radiology 2013-03, Vol.23 (3), p.597-606
Main Authors: Schuhbaeck, Annika, Achenbach, Stephan, Layritz, Christian, Eisentopf, Jasmin, Hecker, Franziska, Pflederer, Tobias, Gauss, Soeren, Rixe, Johannes, Kalender, Willi, Daniel, Werner G., Lell, Michael, Ropers, Dieter
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Objectives We evaluated the potential of prospectively ECG-triggered high-pitch spiral acquisition with low tube voltage and current in combination with iterative reconstruction to achieve coronary CT angiography with sufficient image quality at an effective dose below 0.1 mSv. Methods Contrast-enhanced coronary dual source CT angiography (2 × 128 × 0.6 mm, 80 kV, 50 mAs) in prospectively ECG-triggered high-pitch spiral acquisition mode was performed in 21 consecutive individuals (body weight
ISSN:0938-7994
1432-1084
DOI:10.1007/s00330-012-2656-2