Loading…

Fully referenced single-comb interferometry using optical sampling by laser-cavity tuning

The correction of setup and laser instabilities in a single-comb interferometric measurement using optical sampling by laser-cavity tuning is investigated. A two-reference solution that allows full correction of the interferogram is presented. The technique is compared to a slightly simpler one-refe...

Full description

Saved in:
Bibliographic Details
Published in:Applied optics (2004) 2013-01, Vol.52 (2), p.248-255
Main Authors: Potvin, Simon, Boudreau, Sylvain, Deschênes, Jean-Daniel, Genest, Jérôme
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The correction of setup and laser instabilities in a single-comb interferometric measurement using optical sampling by laser-cavity tuning is investigated. A two-reference solution that allows full correction of the interferogram is presented. The technique is compared to a slightly simpler one-reference correction. For the one-reference case, all the subtleties involved in this partial correction and the dependence between the achievable measurement accuracy and the setup parameters are highlighted. The parameters considered are the comb bandwidth, the laser-frequency noise, the required spectral resolution, the cavity scan speed, and the length of the delay line. For both referencing approaches, experimental results using a fiber delay line of 10 km and a 100 MHz mode-locked laser with its repetition rate swept at 500 Hz are shown.
ISSN:1559-128X
2155-3165
DOI:10.1364/AO.52.000248