Loading…
Separation system suitability (3S): a new criterion of chromatogram classification in HPLC based on cross-evaluation of separation capacity/peak symmetry and its application to complex mixtures of anthraquinones
A comparison of chromatograms obtained in a series of separation conditions for a given complex mixture may be done with a series of chromatographic descriptors. In this study, we used two descriptors: the number of critical pairs and symmetry of peaks, further rescaled and converted to the correspo...
Saved in:
Published in: | Analyst (London) 2013-10, Vol.138 (19), p.581-581 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A comparison of chromatograms obtained in a series of separation conditions for a given complex mixture may be done with a series of chromatographic descriptors. In this study, we used two descriptors: the number of critical pairs and symmetry of peaks, further rescaled and converted to the corresponding critical pairs' coefficient (CP
c
) and symmetry coefficient (S
c
). Considering the difficulty of appreciating global separation quality using CP
c
and S
c
criteria separately, as their respective values are usually uncorrelated, a double-criteria cross-evaluation system was required. For that purpose we tested the commonly used multi-criteria decision-making method - Derringer's desirability function (
D
) - as well as the recently introduced sum of ranking differences (SRD). To facilitate the graphical comparison of both approaches, the desirability function (
D
) was used in the inverse form (
D
inv
). The advantages and drawbacks of both evaluation methods, especially the respective under- or over-evaluation of outliers, caused us to introduce a new ranking approach, separation system suitability (3S). The obtained suitability rankings for the three tested approaches (
D
inv
, SRD and 3S) are different; nevertheless, 3S appears to be the most balanced and the easiest to interpret as well. The approach developed for selection of suitable systems was applied to the problem of separation of complex mixtures through the analysis of a series of standards of anthraquinone derivatives. To judge the pertinence of this evaluation, a sample containing a number of natural anthraquinones extracted from the bark of Indian mulberry (
Morinda citrifolia
) was analysed. In conclusion, the proposed methodology for the cross-evaluation of the series of chromatograms using single specific descriptors (CP
c
and S
c
) through a global composite descriptor (3S) significantly simplifies the decision that separation systems are the most suitable for the separation of complex target mixtures of compounds.
A new global classification approach the separation system suitability (3S) allows improved ranking of thirty-five RPLC separation systems for the analysis of anthraquinoids. |
---|---|
ISSN: | 0003-2654 1364-5528 |
DOI: | 10.1039/c3an00745f |