Loading…
Design and optimization of the readout system for X-ray CCDs
A readout system for X-ray CCDs based on an improved architecture is presented; by optimizing several critical circuit blocks along the analog signal chain, the conflict between the readout speed and readout noise is greatly alleviated. Using CCD47-10 as its target CCD, the readout system has achiev...
Saved in:
Published in: | Chinese physics C 2012-09, Vol.36 (9), p.846-850 |
---|---|
Main Author: | |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A readout system for X-ray CCDs based on an improved architecture is presented; by optimizing several critical circuit blocks along the analog signal chain, the conflict between the readout speed and readout noise is greatly alleviated. Using CCD47-10 as its target CCD, the readout system has achieved 8.6e^- readout noise and 142 eV FWHM at 5.9 keV Mn Kα under a pixel rate of 80 kHz. Also its performance of imaging has been investigated. |
---|---|
ISSN: | 1674-1137 0254-3052 |
DOI: | 10.1088/1674-1137/36/9/009 |