Background Combined post-traumatic aniridia and aphakia demand extensive and complex reconstructive surgery. We present our approach for simultaneous correction of this surgical situation with the use of the ArtificialIris (Dr. Schmidt Intraocularlinsen GmbH, Germany) with a foldable acrylic IOL Len...
Saved in:
Published in: | Graefe's archive for clinical and experimental ophthalmology 2013-03, Vol.251 (3), p.667-675 |
---|---|
Main Authors: | , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Background
Combined post-traumatic aniridia and aphakia demand extensive and complex reconstructive surgery. We present our approach for simultaneous correction of this surgical situation with the use of the ArtificialIris (Dr. Schmidt Intraocularlinsen GmbH, Germany) with a foldable acrylic IOL Lentis L-313 (Oculentis, GmbH, Germany) sutured to its surface. The novelty (our first operation was on June 2010) of this surgical technique is based on the combined use of foldable (with closed haptics) IOL and Artificialiris to correct post-traumatic aniridia and aphakia.
Methods
Four consecutive cases of combined post-traumatic lesions of iris and lens, corrected with complex device ArtificialIris and foldable IOL. In two cases, the compound implant was sutured to the sclera in sulcus during the penetrating keratoplasty; in another case, it was positioned through a corneal incision of about 5.0 mm with transscleral fixation, and in one patient with preserved capsular support and possibility of IOL in-the-bag implantation the ArtificialIris was placed in sulcus sutureless through a clear corneal tunnel.
Results
Maximal follow-up was 6 months. The complex device was placed firmly fixed within the sulcus, including in the eye implanted without sutures, and showed a stable and centered position without any tilt or torque.
Conclusion
Management of post-traumatic aniridia combined with aphakia by haptic fixation of a foldable acrylic IOL on a foldable iris prosthesis appears to be a promising approach which gives the surgeon the possibility to correct a complex lesion with one procedure, which is less traumatic and faster. Existence of foldable materials, both iris and IOL, permits relatively small corneal incisions (4.0–5.0 mm). Moreover, the custom-tailored iris prosthesis gives a perfect aesthetic result. |
---|---|