Loading…

Compact, single-tube scanning tunneling microscope with thermoelectric cooling

We have designed and built a scanning tunneling microscope with a compact inertial-approach mechanism that fits inside the piezoelectric scanner tube. Rigid construction allows the microscope to be operated without the use of external vibration isolators or acoustic enclosures. Thermoelectric coolin...

Full description

Saved in:
Bibliographic Details
Published in:Review of scientific instruments 2013-10, Vol.84 (10), p.103708-103708
Main Authors: Jobbins, Matthew M, Agostino, Christopher J, Michel, Jolai D, Gans, Ashley R, Kandel, S Alex
Format: Article
Language:English
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:We have designed and built a scanning tunneling microscope with a compact inertial-approach mechanism that fits inside the piezoelectric scanner tube. Rigid construction allows the microscope to be operated without the use of external vibration isolators or acoustic enclosures. Thermoelectric cooling and a water-ice bath are used to increase temperature stability when scanning under ambient conditions.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.4825341