Assessment of the integrity of structural shielding of four computed tomography facilities in the greater Accra region of Ghana

The structural shielding thicknesses of the walls of four computed tomography (CT) facilities in Ghana were re-evaluated to verify the shielding integrity using the new shielding design methods recommended by the National Council on Radiological Protection and Measurements (NCRP). The shielding thic...

Full description

Saved in:
Bibliographic Details
Published in:Radiation protection dosimetry 2013-08, Vol.155 (4), p.423-431
Main Authors: Nkansah, A, Schandorf, C, Boadu, M, Fletcher, J J
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The structural shielding thicknesses of the walls of four computed tomography (CT) facilities in Ghana were re-evaluated to verify the shielding integrity using the new shielding design methods recommended by the National Council on Radiological Protection and Measurements (NCRP). The shielding thickness obtained ranged from 120 to 155 mm using default DLP values proposed by the European Commission and 110 to 168 mm using derived DLP values from the four CT manufacturers. These values are within the accepted standard concrete wall thickness ranging from 102 to 152 mm prescribed by the NCRP. The ultrasonic pulse testing of all walls indicated that these are of good quality and free of voids since pulse velocities estimated were within the range of 3.496±0.005 km s(-1). An average dose equivalent rate estimated for supervised areas is 3.4±0.27 µSv week(-1) and that for the controlled area is 18.0±0.15 µSv week(-1), which are within acceptable values.
ISSN:0144-8420
1742-3406
DOI:10.1093/rpd/nct021