Leakage of sevoflurane from vaporizer detected by air monitoring system: a case report

Leakage of inhalational anesthetic gases is an important issue not only to staff health but awareness during general anesthesia. After inhalation of sevoflurane commenced, the audible and visible alarm of our custom-made air monitoring system was activated. The system measured 2–4 ppm of sevoflurane...

Full description

Saved in:
Bibliographic Details
Published in:Journal of clinical monitoring and computing 2013-12, Vol.27 (6), p.629-631
Main Authors: Tanaka, Maiko, Yamaura, Ken, Higashi, Midoriko, Hoka, Sumio
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Leakage of inhalational anesthetic gases is an important issue not only to staff health but awareness during general anesthesia. After inhalation of sevoflurane commenced, the audible and visible alarm of our custom-made air monitoring system was activated. The system measured 2–4 ppm of sevoflurane. A thorough search failed to identify the source of the leak. After changing inhalational anesthesia to intravenous anesthesia, the concentration of sevoflurane decreased immediately to zero. The filling level on the vaporizer decreased faster than usual. A thorough check of the vaporizer after surgery identified a worn down seal in the filling device of the vaporizer.