Heat profiling of phacoemulsification tip using a thermal scanning camera
An experimental study to measure the heat profile of the phacoemulsification (phaco) tip using standard continuous phaco and hyperpulse phaco with and without waveform power modulation in the Millennium Microsurgical System with Custom Control Software (CCS). The phaco tip was imaged in air using a...
Saved in:
Published in: | International ophthalmology 2013-12, Vol.33 (6), p.645-649 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | An experimental study to measure the heat profile of the phacoemulsification (phaco) tip using standard continuous phaco and hyperpulse phaco with and without waveform power modulation in the Millennium Microsurgical System with Custom Control Software (CCS). The phaco tip was imaged in air using a thermal camera. The highest temperature was measured 15 s after application of phaco power. Continuous, hyperpulse and waveform power modulations of the Millennium Microsurgical System were used with different power settings (20, 50 and 100 %) and duty cycles (40, 60 and 90 %), with the irrigation turned on and off. Using continuous phaco with the irrigation on, the phaco tip temperature remains |
---|---|
ISSN: | 0165-5701 1573-2630 |
DOI: | 10.1007/s10792-013-9752-4 |