Loading…
Study on the imaging ability of the 2D neutron detector based on MWPC
A 2D neutron detector based on 3He convertor and MWPC with an active area of 200 mmx200 mm has been successfully designed and fabricated. The detector has been tested with Am/Be neutron source and with collimated neutron beam with the wavelength of λ=1.37A. The best spatial resolution of 1.18 mm (FW...
Saved in:
Published in: | Science China. Physics, mechanics & astronomy mechanics & astronomy, 2013-10, Vol.56 (10), p.1882-1886 |
---|---|
Main Authors: | , , , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A 2D neutron detector based on 3He convertor and MWPC with an active area of 200 mmx200 mm has been successfully designed and fabricated. The detector has been tested with Am/Be neutron source and with collimated neutron beam with the wavelength of λ=1.37A. The best spatial resolution of 1.18 mm (FWHM) and good linearity were obtained. This is in good agreement with theoretical calculations. |
---|---|
ISSN: | 1674-7348 1869-1927 |
DOI: | 10.1007/s11433-013-5187-1 |