Loading…
In Vivo Analysis of Insertional Torque During Pedicle Screwing Using Cortical Bone Trajectory Technique
STUDY DESIGN.The insertional torque of pedicle screws using the cortical bone trajectory (CBT) was measured in vivo. OBJECTIVE.To investigate the effectiveness of the CBT technique by measurement of the insertional torque. SUMMARY OF BACKGROUND DATA.The CBT follows a mediolateral and caudocephalad d...
Saved in:
Published in: | Spine (Philadelphia, Pa. 1976) Pa. 1976), 2014-02, Vol.39 (4), p.E240-E245 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | STUDY DESIGN.The insertional torque of pedicle screws using the cortical bone trajectory (CBT) was measured in vivo.
OBJECTIVE.To investigate the effectiveness of the CBT technique by measurement of the insertional torque.
SUMMARY OF BACKGROUND DATA.The CBT follows a mediolateral and caudocephalad directed path, engaging with cortical bone maximally from the pedicle to the vertebral body. Some biomechanical studies have demonstrated favorable characteristics of the CBT technique in cadaveric lumbar spine. However, no in vivo study has been reported on the mechanical behavior of this new trajectory.
METHODS.The insertional torque of pedicle screws using CBT and traditional techniques were measured intraoperatively in 48 consecutive patients. A total of 162 screws using the CBT technique and 36 screws using the traditional technique were compared. In 8 of 48 patients, the side-by-side comparison of 2 different insertional techniques for each vertebra were performed, which formed the H group. In addition, the insertional torque was correlated with bone mineral density.
RESULTS.The mean maximum insertional torque of CBT screws and traditional screws were 2.49 ± 0.99 Nm and 1.24 ± 0.54 Nm, respectively. The CBT screws showed 2.01 times higher torque and the difference was significant between the 2 techniques (P < 0.01). In the H group, the insertional torque were 2.71 ± 1.36 Nm in the CBT screws and 1.58 ± 0.44 Nm in the traditional screws. The CBT screws demonstrated 1.71 times higher torque and statistical significance was achieved (P < 0.01). Positive linear correlations between maximum insertional torque and bone mineral density were found in both technique, the correlation coefficient of traditional screws (r = 0.63, P < 0.01) was higher than that of the CBT screws (r = 0.59, P < 0.01).
CONCLUSION.The insertional torque using the CBT technique is about 1.7 times higher than the traditional technique.Level of Evidence2 |
---|---|
ISSN: | 0362-2436 1528-1159 |
DOI: | 10.1097/BRS.0000000000000116 |