Loading…
Study to Eliminate the Effect of Hyperbilirubinemia in Measurement of Blood Coagulation Assay
In this paper, a low cost coagulation analyzer is developed using an LPC 2138 Microcontroller to determine the values of PT (Prothrombin Time) and APTT (Activated partial Thomboplastin Time) in samples with Hyperbilirubinimea (High Bilirubin) ranging from 0-22.0 mg/dl. Most of the conventional Instr...
Saved in:
Published in: | Sensors & transducers 2013-02, Vol.149 (2), p.187-187 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | In this paper, a low cost coagulation analyzer is developed using an LPC 2138 Microcontroller to determine the values of PT (Prothrombin Time) and APTT (Activated partial Thomboplastin Time) in samples with Hyperbilirubinimea (High Bilirubin) ranging from 0-22.0 mg/dl. Most of the conventional Instruments are able to detect the clotting time only when the bilirubin level in the sample is within the range of 0-5.0 mg/dl. In the developed instrument yellow light (LED 590 nm) having a minimum absorbance (Beer-Lambert's Law) while passing through the yellowish sample is used to detect the changes in light intensity while clotting of the sample occurs and thus eliminates the interference of higher bilirubin content (above 4 mg/dl) in blood samples. It is observed that the designed Instrument shows the performance of measurement of PT and APTT in 95.5% samples with hyperbilirubinemia against 31.8% of samples in COAT-1 analyzer. |
---|---|
ISSN: | 2306-8515 1726-5479 1726-5479 |