Estimation of distance error by fuzzy set theory required for strength determination of HDR (192)Ir brachytherapy sources

Verification of the strength of high dose rate (HDR) (192)Ir brachytherapy sources on receipt from the vendor is an important component of institutional quality assurance program. Either reference air-kerma rate (RAKR) or air-kerma strength (AKS) is the recommended quantity to specify the strength o...

Full description

Saved in:
Bibliographic Details
Published in:Journal of medical physics 2014-04, Vol.39 (2), p.85-92
Main Authors: Kumar, Sudhir, Datta, D, Sharma, S D, Chourasiya, G, Babu, D A R, Sharma, D N
Format: Article
Language:English
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by
cites
container_end_page 92
container_issue 2
container_start_page 85
container_title Journal of medical physics
container_volume 39
creator Kumar, Sudhir
Datta, D
Sharma, S D
Chourasiya, G
Babu, D A R
Sharma, D N
description Verification of the strength of high dose rate (HDR) (192)Ir brachytherapy sources on receipt from the vendor is an important component of institutional quality assurance program. Either reference air-kerma rate (RAKR) or air-kerma strength (AKS) is the recommended quantity to specify the strength of gamma-emitting brachytherapy sources. The use of Farmer-type cylindrical ionization chamber of sensitive volume 0.6 cm(3) is one of the recommended methods for measuring RAKR of HDR (192)Ir brachytherapy sources. While using the cylindrical chamber method, it is required to determine the positioning error of the ionization chamber with respect to the source which is called the distance error. An attempt has been made to apply the fuzzy set theory to estimate the subjective uncertainty associated with the distance error. A simplified approach of applying this fuzzy set theory has been proposed in the quantification of uncertainty associated with the distance error. In order to express the uncertainty in the framework of fuzzy sets, the uncertainty index was estimated and was found to be within 2.5%, which further indicates that the possibility of error in measuring such distance may be of this order. It is observed that the relative distance li estimated by analytical method and fuzzy set theoretic approach are consistent with each other. The crisp values of li estimated using analytical method lie within the bounds computed using fuzzy set theory. This indicates that li values estimated using analytical methods are within 2.5% uncertainty. This value of uncertainty in distance measurement should be incorporated in the uncertainty budget, while estimating the expanded uncertainty in HDR (192)Ir source strength measurement.
doi_str_mv 10.4103/0971-6203.131281
format article
fullrecord <record><control><sourceid>proquest_pubme</sourceid><recordid>TN_cdi_proquest_miscellaneous_1530956726</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>1530956726</sourcerecordid><originalsourceid>FETCH-LOGICAL-p126t-50a6201607f6e5b839f008559d5aa391b10e9b5b143288bda49765a4abce02db3</originalsourceid><addsrcrecordid>eNo9kD1PwzAQQD2AaCnsTMhjGVL8nXhEpdBKlZAQzJGdXGhQ81HbGdJfX0uUTied3j3pHkIPlCwEJfyZ6JQmihG-oJyyjF6h6WU1Qbfe_xIipBbiBk2YyFKmiJyiceVD3ZhQdy3uKlzWPpi2AAzOdQ7bEVfD8ThiDwGHHXRuxA4OQ-2gxFUEfHDQ_oQdLiGAa-r2Ylq_fuI51expEzXOFLsx3jvTR1c3uAL8HbquzN7D_XnO0Pfb6mu5TrYf75vlyzbpKVMhkcTED6giaaVA2ozripBMSl1KY7imlhLQVloqOMsyWxqhUyWNMLYAwkrLZ2j-5-1ddxjAh7ypfQH7vWmhG3xOJSdaqtgjoo9ndLANlHnvYho35v-5-AmU5WtF</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1530956726</pqid></control><display><type>article</type><title>Estimation of distance error by fuzzy set theory required for strength determination of HDR (192)Ir brachytherapy sources</title><source>Publicly Available Content Database (Proquest) (PQ_SDU_P3)</source><source>PubMed Central (Open access)</source><source>Medknow Open Access Medical Journals</source><source>IngentaConnect Journals</source><creator>Kumar, Sudhir ; Datta, D ; Sharma, S D ; Chourasiya, G ; Babu, D A R ; Sharma, D N</creator><creatorcontrib>Kumar, Sudhir ; Datta, D ; Sharma, S D ; Chourasiya, G ; Babu, D A R ; Sharma, D N</creatorcontrib><description>Verification of the strength of high dose rate (HDR) (192)Ir brachytherapy sources on receipt from the vendor is an important component of institutional quality assurance program. Either reference air-kerma rate (RAKR) or air-kerma strength (AKS) is the recommended quantity to specify the strength of gamma-emitting brachytherapy sources. The use of Farmer-type cylindrical ionization chamber of sensitive volume 0.6 cm(3) is one of the recommended methods for measuring RAKR of HDR (192)Ir brachytherapy sources. While using the cylindrical chamber method, it is required to determine the positioning error of the ionization chamber with respect to the source which is called the distance error. An attempt has been made to apply the fuzzy set theory to estimate the subjective uncertainty associated with the distance error. A simplified approach of applying this fuzzy set theory has been proposed in the quantification of uncertainty associated with the distance error. In order to express the uncertainty in the framework of fuzzy sets, the uncertainty index was estimated and was found to be within 2.5%, which further indicates that the possibility of error in measuring such distance may be of this order. It is observed that the relative distance li estimated by analytical method and fuzzy set theoretic approach are consistent with each other. The crisp values of li estimated using analytical method lie within the bounds computed using fuzzy set theory. This indicates that li values estimated using analytical methods are within 2.5% uncertainty. This value of uncertainty in distance measurement should be incorporated in the uncertainty budget, while estimating the expanded uncertainty in HDR (192)Ir source strength measurement.</description><identifier>ISSN: 0971-6203</identifier><identifier>DOI: 10.4103/0971-6203.131281</identifier><identifier>PMID: 24872605</identifier><language>eng</language><publisher>India</publisher><ispartof>Journal of medical physics, 2014-04, Vol.39 (2), p.85-92</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27924,27925,37013</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/24872605$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Kumar, Sudhir</creatorcontrib><creatorcontrib>Datta, D</creatorcontrib><creatorcontrib>Sharma, S D</creatorcontrib><creatorcontrib>Chourasiya, G</creatorcontrib><creatorcontrib>Babu, D A R</creatorcontrib><creatorcontrib>Sharma, D N</creatorcontrib><title>Estimation of distance error by fuzzy set theory required for strength determination of HDR (192)Ir brachytherapy sources</title><title>Journal of medical physics</title><addtitle>J Med Phys</addtitle><description>Verification of the strength of high dose rate (HDR) (192)Ir brachytherapy sources on receipt from the vendor is an important component of institutional quality assurance program. Either reference air-kerma rate (RAKR) or air-kerma strength (AKS) is the recommended quantity to specify the strength of gamma-emitting brachytherapy sources. The use of Farmer-type cylindrical ionization chamber of sensitive volume 0.6 cm(3) is one of the recommended methods for measuring RAKR of HDR (192)Ir brachytherapy sources. While using the cylindrical chamber method, it is required to determine the positioning error of the ionization chamber with respect to the source which is called the distance error. An attempt has been made to apply the fuzzy set theory to estimate the subjective uncertainty associated with the distance error. A simplified approach of applying this fuzzy set theory has been proposed in the quantification of uncertainty associated with the distance error. In order to express the uncertainty in the framework of fuzzy sets, the uncertainty index was estimated and was found to be within 2.5%, which further indicates that the possibility of error in measuring such distance may be of this order. It is observed that the relative distance li estimated by analytical method and fuzzy set theoretic approach are consistent with each other. The crisp values of li estimated using analytical method lie within the bounds computed using fuzzy set theory. This indicates that li values estimated using analytical methods are within 2.5% uncertainty. This value of uncertainty in distance measurement should be incorporated in the uncertainty budget, while estimating the expanded uncertainty in HDR (192)Ir source strength measurement.</description><issn>0971-6203</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2014</creationdate><recordtype>article</recordtype><recordid>eNo9kD1PwzAQQD2AaCnsTMhjGVL8nXhEpdBKlZAQzJGdXGhQ81HbGdJfX0uUTied3j3pHkIPlCwEJfyZ6JQmihG-oJyyjF6h6WU1Qbfe_xIipBbiBk2YyFKmiJyiceVD3ZhQdy3uKlzWPpi2AAzOdQ7bEVfD8ThiDwGHHXRuxA4OQ-2gxFUEfHDQ_oQdLiGAa-r2Ylq_fuI51expEzXOFLsx3jvTR1c3uAL8HbquzN7D_XnO0Pfb6mu5TrYf75vlyzbpKVMhkcTED6giaaVA2ozripBMSl1KY7imlhLQVloqOMsyWxqhUyWNMLYAwkrLZ2j-5-1ddxjAh7ypfQH7vWmhG3xOJSdaqtgjoo9ndLANlHnvYho35v-5-AmU5WtF</recordid><startdate>201404</startdate><enddate>201404</enddate><creator>Kumar, Sudhir</creator><creator>Datta, D</creator><creator>Sharma, S D</creator><creator>Chourasiya, G</creator><creator>Babu, D A R</creator><creator>Sharma, D N</creator><scope>NPM</scope><scope>7X8</scope></search><sort><creationdate>201404</creationdate><title>Estimation of distance error by fuzzy set theory required for strength determination of HDR (192)Ir brachytherapy sources</title><author>Kumar, Sudhir ; Datta, D ; Sharma, S D ; Chourasiya, G ; Babu, D A R ; Sharma, D N</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-p126t-50a6201607f6e5b839f008559d5aa391b10e9b5b143288bda49765a4abce02db3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2014</creationdate><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Kumar, Sudhir</creatorcontrib><creatorcontrib>Datta, D</creatorcontrib><creatorcontrib>Sharma, S D</creatorcontrib><creatorcontrib>Chourasiya, G</creatorcontrib><creatorcontrib>Babu, D A R</creatorcontrib><creatorcontrib>Sharma, D N</creatorcontrib><collection>PubMed</collection><collection>MEDLINE - Academic</collection><jtitle>Journal of medical physics</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Kumar, Sudhir</au><au>Datta, D</au><au>Sharma, S D</au><au>Chourasiya, G</au><au>Babu, D A R</au><au>Sharma, D N</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Estimation of distance error by fuzzy set theory required for strength determination of HDR (192)Ir brachytherapy sources</atitle><jtitle>Journal of medical physics</jtitle><addtitle>J Med Phys</addtitle><date>2014-04</date><risdate>2014</risdate><volume>39</volume><issue>2</issue><spage>85</spage><epage>92</epage><pages>85-92</pages><issn>0971-6203</issn><abstract>Verification of the strength of high dose rate (HDR) (192)Ir brachytherapy sources on receipt from the vendor is an important component of institutional quality assurance program. Either reference air-kerma rate (RAKR) or air-kerma strength (AKS) is the recommended quantity to specify the strength of gamma-emitting brachytherapy sources. The use of Farmer-type cylindrical ionization chamber of sensitive volume 0.6 cm(3) is one of the recommended methods for measuring RAKR of HDR (192)Ir brachytherapy sources. While using the cylindrical chamber method, it is required to determine the positioning error of the ionization chamber with respect to the source which is called the distance error. An attempt has been made to apply the fuzzy set theory to estimate the subjective uncertainty associated with the distance error. A simplified approach of applying this fuzzy set theory has been proposed in the quantification of uncertainty associated with the distance error. In order to express the uncertainty in the framework of fuzzy sets, the uncertainty index was estimated and was found to be within 2.5%, which further indicates that the possibility of error in measuring such distance may be of this order. It is observed that the relative distance li estimated by analytical method and fuzzy set theoretic approach are consistent with each other. The crisp values of li estimated using analytical method lie within the bounds computed using fuzzy set theory. This indicates that li values estimated using analytical methods are within 2.5% uncertainty. This value of uncertainty in distance measurement should be incorporated in the uncertainty budget, while estimating the expanded uncertainty in HDR (192)Ir source strength measurement.</abstract><cop>India</cop><pmid>24872605</pmid><doi>10.4103/0971-6203.131281</doi><tpages>8</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0971-6203
ispartof Journal of medical physics, 2014-04, Vol.39 (2), p.85-92
issn 0971-6203
language eng
recordid cdi_proquest_miscellaneous_1530956726
source Publicly Available Content Database (Proquest) (PQ_SDU_P3); PubMed Central (Open access); Medknow Open Access Medical Journals; IngentaConnect Journals
title Estimation of distance error by fuzzy set theory required for strength determination of HDR (192)Ir brachytherapy sources
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-06T20%3A38%3A32IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Estimation%20of%20distance%20error%20by%20fuzzy%20set%20theory%20required%20for%20strength%20determination%20of%20HDR%20(192)Ir%20brachytherapy%20sources&rft.jtitle=Journal%20of%20medical%20physics&rft.au=Kumar,%20Sudhir&rft.date=2014-04&rft.volume=39&rft.issue=2&rft.spage=85&rft.epage=92&rft.pages=85-92&rft.issn=0971-6203&rft_id=info:doi/10.4103/0971-6203.131281&rft_dat=%3Cproquest_pubme%3E1530956726%3C/proquest_pubme%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-p126t-50a6201607f6e5b839f008559d5aa391b10e9b5b143288bda49765a4abce02db3%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=1530956726&rft_id=info:pmid/24872605&rfr_iscdi=true