Loading…
Simultaneous determination of Cs and Sr in the high level waste by ion-chromatography
A method based on ion-chromatography has been developed for determination of Cs and Sr in high level waste (HLW) without matrix separation. The acidity of HLW (3 M HNO 3 ) was decreased to 0.01 M by addition of NaOH. Chromatographic separation of Cs + and Sr ++ in presence of bulk Na + has achieved...
Saved in:
Published in: | Journal of radioanalytical and nuclear chemistry 2014-04, Vol.300 (1), p.121-124 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A method based on ion-chromatography has been developed for determination of Cs and Sr in high level waste (HLW) without matrix separation. The acidity of HLW (3 M HNO
3
) was decreased to 0.01 M by addition of NaOH. Chromatographic separation of Cs
+
and Sr
++
in presence of bulk Na
+
has achieved using methane sulphonic acid as mobile phase in isocratic and gradient mode. The calibration plot was linear for the concentration range of 0.5–12 mg/L for Cs and 0.025–6 mg/L for Sr with regression coefficients close to 1. RSD obtained for Cs and Sr was 1 and 5 % respectively. Detection limit calculated as 3(S/N) was found to be 20 μg/L for Cs and 30 μg/L for Sr. The standard addition procedure was used to validate the developed method. The samples received from PREFRE, Tarapur and WIP, BARC were analyzed by the new method and the results are presented. |
---|---|
ISSN: | 0236-5731 1588-2780 |
DOI: | 10.1007/s10967-014-2954-z |