A technique to measure the intracuff pressure continuously: an in vivo demonstration of its accuracy
Summary Objective A major concern with the use of cuffed endotracheal tubes (cETT) in children is hyperinflation of the cuff which may compromise tracheal mucosal perfusion. To measure the intracuff pressure (CP), we devised a method using the transducer of an invasive pressure monitoring device. Th...
Saved in:
Published in: | Pediatric anesthesia 2014-09, Vol.24 (9), p.999-1004 |
---|---|
Main Authors: | , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Summary
Objective
A major concern with the use of cuffed endotracheal tubes (cETT) in children is hyperinflation of the cuff which may compromise tracheal mucosal perfusion. To measure the intracuff pressure (CP), we devised a method using the transducer of an invasive pressure monitoring device. The objective of the study was to test the accuracy and validity of this device for instantaneous and continuous CP monitoring.
Methods
The study was conducted in 2 phases. In Phase 1 (200 pediatric patients), after inflation of the cuff, the CP was measured using the standard manometer and the transducer simultaneously. In Phase 2 (20 pediatric patients), the transducer was left connected to the pilot balloon of the ETT to obtain a continuous CP reading and the standard manometer was used to measure the CP at 5‐min intervals. Statistical analysis included a Bland–Altman comparison and linear regression analysis.
Results
In Phase 1, linear regression analysis demonstrated an R2 value of 0.9956. The bias was 0.30 cmH2O, the precision was 0.75 cmH2O, and the 95% level of agreement (LOA) ranged from −1.16 to 1.77 cmH2O. In Phase 2, the linear regression analysis revealed an R2 value of 0.9846. The bias was 0.28 cmH2O, the precision was 0.7 cmH2O, and the 95% LOA ranged from −1.1 to 1.66 cmH2O.
Conclusion
Our study demonstrates that when cETTs are used in the pediatric population, the transducer of the invasive pressure monitoring device can be used reliably to measure the CP at the time of inflation and continuously thereafter. |
---|---|