Assessment of the effective dose in supine, prone, and oblique positions in the maxillofacial region using a novel combined extremity and maxillofacial cone beam computed tomography scanner

Objective The objectives of this study were to assess the organ and effective doses (International Commission on Radiological Protection [ICRP] 103 standard) resulting from supine, prone, and oblique phantom positions in the maxillofacial region using a novel cone beam computed tomography (CBCT) dev...

Full description

Saved in:
Bibliographic Details
Published in:Oral surgery, oral medicine, oral pathology and oral radiology oral medicine, oral pathology and oral radiology, 2014-09, Vol.118 (3), p.355-362
Main Authors: Koivisto, J., PhLic, Wolff, J., DMD, Järnstedt, J., DDS, Dastidar, P., MD, PhD, Kortesniemi, M., PhD
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by cdi_FETCH-LOGICAL-c411t-27487bedfec02d6e01c517d2830395c374d75fda8162764c279d4d771f92c87e3
cites cdi_FETCH-LOGICAL-c411t-27487bedfec02d6e01c517d2830395c374d75fda8162764c279d4d771f92c87e3
container_end_page 362
container_issue 3
container_start_page 355
container_title Oral surgery, oral medicine, oral pathology and oral radiology
container_volume 118
creator Koivisto, J., PhLic
Wolff, J., DMD
Järnstedt, J., DDS
Dastidar, P., MD, PhD
Kortesniemi, M., PhD
description Objective The objectives of this study were to assess the organ and effective doses (International Commission on Radiological Protection [ICRP] 103 standard) resulting from supine, prone, and oblique phantom positions in the maxillofacial region using a novel cone beam computed tomography (CBCT) device and to compare the results with conventional dental CBCT and multislice computed tomography (MSCT) devices. Study Design Measurements were performed using an anthropomorphic RANDO head phantom (Radiation Analogue Dosimetry System) with 20 MOSFET (metal-oxide-semiconductor field-effect transistor) dosimeters placed in the most radiosensitive maxillofacial organs. Effective doses were measured in 3 phantom positions using a combined extremity and maxillofacial CBCT device: the Planmed Verity CBCT scanner. Reference values were measured in the upright position with 2 CBCT devices and in the prone position with one MSCT scanner. Results The Planmed Verity CBCT scanner effective doses were 247 μSv in supine, 192 μSv in prone, and 134 μSv in oblique position. The effective dose with ProMax 3D MAX CBCT was 168 μSv; with i-CAT Next Generation, 170 μSv; and with Philips Brilliance 64 MSCT, 781 μSv. Conclusions Head positioning has an important effect on the organ and effective doses. The Planmed Verity CBCT scanner effective dose results were comparable with those attained on 2 conventional CBCT devices and were considerably lower than the MSCT scanner results.
doi_str_mv 10.1016/j.oooo.2014.05.016
format article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_1556290011</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S2212440314005124</els_id><sourcerecordid>1556290011</sourcerecordid><originalsourceid>FETCH-LOGICAL-c411t-27487bedfec02d6e01c517d2830395c374d75fda8162764c279d4d771f92c87e3</originalsourceid><addsrcrecordid>eNp9Uk2LFDEQbURxl3X_gAfJ0YPTJumvaRBhWfyCBQ_qOaST6tmM6aRNpYedH-d_s9pZF_Rgcqii8t4LVa-K4rngpeCifb0vI51SclGXvCmp9Kg4l1LITV0L8fgh59VZcYm453RaQtXyaXEmG0G35-fFzytEQJwgZBZHlm-BwTiCye4AzEYE5gLDZXYBXrE5xTXoYFkcvPuxAJsjuuxiwBW3sid957yPozZOe5ZgR49sQRd2TLMQD-CZidNAepbBXU4wuXz8Lfk309BXbAA9rfB5yQTPcYq7pOfbI0OjQ4D0rHgyao9weR8vim_v3329_ri5-fzh0_XVzcbQLPJGdvW2G8BSX1zaFrgwjeis3Fa86htTdbXtmtHqrWhl19ZGdr2lUifGXpptB9VF8fKkSxOgrjGryaEB73WAuKASTdPKnnMhCCpPUJMiYoJRzclNOh2V4Gp1Tu3V6pxanVO8UVQi0ot7_WWYwD5Q_vhEgDcnAFCXBwdJoXEQDFiXyCxlo_u__tt_6Ma74Iz23-EIuI9LCjQ_JRRKxdWXdXfW1RE15w2l1S8S6cK7</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1556290011</pqid></control><display><type>article</type><title>Assessment of the effective dose in supine, prone, and oblique positions in the maxillofacial region using a novel combined extremity and maxillofacial cone beam computed tomography scanner</title><source>ScienceDirect Freedom Collection</source><creator>Koivisto, J., PhLic ; Wolff, J., DMD ; Järnstedt, J., DDS ; Dastidar, P., MD, PhD ; Kortesniemi, M., PhD</creator><creatorcontrib>Koivisto, J., PhLic ; Wolff, J., DMD ; Järnstedt, J., DDS ; Dastidar, P., MD, PhD ; Kortesniemi, M., PhD</creatorcontrib><description>Objective The objectives of this study were to assess the organ and effective doses (International Commission on Radiological Protection [ICRP] 103 standard) resulting from supine, prone, and oblique phantom positions in the maxillofacial region using a novel cone beam computed tomography (CBCT) device and to compare the results with conventional dental CBCT and multislice computed tomography (MSCT) devices. Study Design Measurements were performed using an anthropomorphic RANDO head phantom (Radiation Analogue Dosimetry System) with 20 MOSFET (metal-oxide-semiconductor field-effect transistor) dosimeters placed in the most radiosensitive maxillofacial organs. Effective doses were measured in 3 phantom positions using a combined extremity and maxillofacial CBCT device: the Planmed Verity CBCT scanner. Reference values were measured in the upright position with 2 CBCT devices and in the prone position with one MSCT scanner. Results The Planmed Verity CBCT scanner effective doses were 247 μSv in supine, 192 μSv in prone, and 134 μSv in oblique position. The effective dose with ProMax 3D MAX CBCT was 168 μSv; with i-CAT Next Generation, 170 μSv; and with Philips Brilliance 64 MSCT, 781 μSv. Conclusions Head positioning has an important effect on the organ and effective doses. The Planmed Verity CBCT scanner effective dose results were comparable with those attained on 2 conventional CBCT devices and were considerably lower than the MSCT scanner results.</description><identifier>ISSN: 2212-4403</identifier><identifier>EISSN: 2212-4411</identifier><identifier>DOI: 10.1016/j.oooo.2014.05.016</identifier><identifier>PMID: 25151590</identifier><language>eng</language><publisher>United States: Elsevier Inc</publisher><subject>Cone-Beam Computed Tomography - instrumentation ; Dentistry ; Head - diagnostic imaging ; Humans ; Patient Positioning ; Phantoms, Imaging ; Prone Position ; Radiation Dosage ; Radiometry - instrumentation ; Supine Position ; Surgery</subject><ispartof>Oral surgery, oral medicine, oral pathology and oral radiology, 2014-09, Vol.118 (3), p.355-362</ispartof><rights>Elsevier Inc.</rights><rights>2014 Elsevier Inc.</rights><rights>Copyright © 2014 Elsevier Inc. All rights reserved.</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c411t-27487bedfec02d6e01c517d2830395c374d75fda8162764c279d4d771f92c87e3</citedby><cites>FETCH-LOGICAL-c411t-27487bedfec02d6e01c517d2830395c374d75fda8162764c279d4d771f92c87e3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,777,781,27905,27906</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/25151590$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Koivisto, J., PhLic</creatorcontrib><creatorcontrib>Wolff, J., DMD</creatorcontrib><creatorcontrib>Järnstedt, J., DDS</creatorcontrib><creatorcontrib>Dastidar, P., MD, PhD</creatorcontrib><creatorcontrib>Kortesniemi, M., PhD</creatorcontrib><title>Assessment of the effective dose in supine, prone, and oblique positions in the maxillofacial region using a novel combined extremity and maxillofacial cone beam computed tomography scanner</title><title>Oral surgery, oral medicine, oral pathology and oral radiology</title><addtitle>Oral Surg Oral Med Oral Pathol Oral Radiol</addtitle><description>Objective The objectives of this study were to assess the organ and effective doses (International Commission on Radiological Protection [ICRP] 103 standard) resulting from supine, prone, and oblique phantom positions in the maxillofacial region using a novel cone beam computed tomography (CBCT) device and to compare the results with conventional dental CBCT and multislice computed tomography (MSCT) devices. Study Design Measurements were performed using an anthropomorphic RANDO head phantom (Radiation Analogue Dosimetry System) with 20 MOSFET (metal-oxide-semiconductor field-effect transistor) dosimeters placed in the most radiosensitive maxillofacial organs. Effective doses were measured in 3 phantom positions using a combined extremity and maxillofacial CBCT device: the Planmed Verity CBCT scanner. Reference values were measured in the upright position with 2 CBCT devices and in the prone position with one MSCT scanner. Results The Planmed Verity CBCT scanner effective doses were 247 μSv in supine, 192 μSv in prone, and 134 μSv in oblique position. The effective dose with ProMax 3D MAX CBCT was 168 μSv; with i-CAT Next Generation, 170 μSv; and with Philips Brilliance 64 MSCT, 781 μSv. Conclusions Head positioning has an important effect on the organ and effective doses. The Planmed Verity CBCT scanner effective dose results were comparable with those attained on 2 conventional CBCT devices and were considerably lower than the MSCT scanner results.</description><subject>Cone-Beam Computed Tomography - instrumentation</subject><subject>Dentistry</subject><subject>Head - diagnostic imaging</subject><subject>Humans</subject><subject>Patient Positioning</subject><subject>Phantoms, Imaging</subject><subject>Prone Position</subject><subject>Radiation Dosage</subject><subject>Radiometry - instrumentation</subject><subject>Supine Position</subject><subject>Surgery</subject><issn>2212-4403</issn><issn>2212-4411</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2014</creationdate><recordtype>article</recordtype><recordid>eNp9Uk2LFDEQbURxl3X_gAfJ0YPTJumvaRBhWfyCBQ_qOaST6tmM6aRNpYedH-d_s9pZF_Rgcqii8t4LVa-K4rngpeCifb0vI51SclGXvCmp9Kg4l1LITV0L8fgh59VZcYm453RaQtXyaXEmG0G35-fFzytEQJwgZBZHlm-BwTiCye4AzEYE5gLDZXYBXrE5xTXoYFkcvPuxAJsjuuxiwBW3sid957yPozZOe5ZgR49sQRd2TLMQD-CZidNAepbBXU4wuXz8Lfk309BXbAA9rfB5yQTPcYq7pOfbI0OjQ4D0rHgyao9weR8vim_v3329_ri5-fzh0_XVzcbQLPJGdvW2G8BSX1zaFrgwjeis3Fa86htTdbXtmtHqrWhl19ZGdr2lUifGXpptB9VF8fKkSxOgrjGryaEB73WAuKASTdPKnnMhCCpPUJMiYoJRzclNOh2V4Gp1Tu3V6pxanVO8UVQi0ot7_WWYwD5Q_vhEgDcnAFCXBwdJoXEQDFiXyCxlo_u__tt_6Ma74Iz23-EIuI9LCjQ_JRRKxdWXdXfW1RE15w2l1S8S6cK7</recordid><startdate>20140901</startdate><enddate>20140901</enddate><creator>Koivisto, J., PhLic</creator><creator>Wolff, J., DMD</creator><creator>Järnstedt, J., DDS</creator><creator>Dastidar, P., MD, PhD</creator><creator>Kortesniemi, M., PhD</creator><general>Elsevier Inc</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>20140901</creationdate><title>Assessment of the effective dose in supine, prone, and oblique positions in the maxillofacial region using a novel combined extremity and maxillofacial cone beam computed tomography scanner</title><author>Koivisto, J., PhLic ; Wolff, J., DMD ; Järnstedt, J., DDS ; Dastidar, P., MD, PhD ; Kortesniemi, M., PhD</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c411t-27487bedfec02d6e01c517d2830395c374d75fda8162764c279d4d771f92c87e3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2014</creationdate><topic>Cone-Beam Computed Tomography - instrumentation</topic><topic>Dentistry</topic><topic>Head - diagnostic imaging</topic><topic>Humans</topic><topic>Patient Positioning</topic><topic>Phantoms, Imaging</topic><topic>Prone Position</topic><topic>Radiation Dosage</topic><topic>Radiometry - instrumentation</topic><topic>Supine Position</topic><topic>Surgery</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Koivisto, J., PhLic</creatorcontrib><creatorcontrib>Wolff, J., DMD</creatorcontrib><creatorcontrib>Järnstedt, J., DDS</creatorcontrib><creatorcontrib>Dastidar, P., MD, PhD</creatorcontrib><creatorcontrib>Kortesniemi, M., PhD</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Oral surgery, oral medicine, oral pathology and oral radiology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Koivisto, J., PhLic</au><au>Wolff, J., DMD</au><au>Järnstedt, J., DDS</au><au>Dastidar, P., MD, PhD</au><au>Kortesniemi, M., PhD</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Assessment of the effective dose in supine, prone, and oblique positions in the maxillofacial region using a novel combined extremity and maxillofacial cone beam computed tomography scanner</atitle><jtitle>Oral surgery, oral medicine, oral pathology and oral radiology</jtitle><addtitle>Oral Surg Oral Med Oral Pathol Oral Radiol</addtitle><date>2014-09-01</date><risdate>2014</risdate><volume>118</volume><issue>3</issue><spage>355</spage><epage>362</epage><pages>355-362</pages><issn>2212-4403</issn><eissn>2212-4411</eissn><abstract>Objective The objectives of this study were to assess the organ and effective doses (International Commission on Radiological Protection [ICRP] 103 standard) resulting from supine, prone, and oblique phantom positions in the maxillofacial region using a novel cone beam computed tomography (CBCT) device and to compare the results with conventional dental CBCT and multislice computed tomography (MSCT) devices. Study Design Measurements were performed using an anthropomorphic RANDO head phantom (Radiation Analogue Dosimetry System) with 20 MOSFET (metal-oxide-semiconductor field-effect transistor) dosimeters placed in the most radiosensitive maxillofacial organs. Effective doses were measured in 3 phantom positions using a combined extremity and maxillofacial CBCT device: the Planmed Verity CBCT scanner. Reference values were measured in the upright position with 2 CBCT devices and in the prone position with one MSCT scanner. Results The Planmed Verity CBCT scanner effective doses were 247 μSv in supine, 192 μSv in prone, and 134 μSv in oblique position. The effective dose with ProMax 3D MAX CBCT was 168 μSv; with i-CAT Next Generation, 170 μSv; and with Philips Brilliance 64 MSCT, 781 μSv. Conclusions Head positioning has an important effect on the organ and effective doses. The Planmed Verity CBCT scanner effective dose results were comparable with those attained on 2 conventional CBCT devices and were considerably lower than the MSCT scanner results.</abstract><cop>United States</cop><pub>Elsevier Inc</pub><pmid>25151590</pmid><doi>10.1016/j.oooo.2014.05.016</doi><tpages>8</tpages></addata></record>
fulltext fulltext
identifier ISSN: 2212-4403
ispartof Oral surgery, oral medicine, oral pathology and oral radiology, 2014-09, Vol.118 (3), p.355-362
issn 2212-4403
2212-4411
language eng
recordid cdi_proquest_miscellaneous_1556290011
source ScienceDirect Freedom Collection
subjects Cone-Beam Computed Tomography - instrumentation
Dentistry
Head - diagnostic imaging
Humans
Patient Positioning
Phantoms, Imaging
Prone Position
Radiation Dosage
Radiometry - instrumentation
Supine Position
Surgery
title Assessment of the effective dose in supine, prone, and oblique positions in the maxillofacial region using a novel combined extremity and maxillofacial cone beam computed tomography scanner
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-21T06%3A32%3A22IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Assessment%20of%20the%20effective%20dose%20in%20supine,%20prone,%20and%20oblique%20positions%20in%20the%20maxillofacial%20region%20using%20a%20novel%20combined%20extremity%20and%20maxillofacial%20cone%20beam%20computed%20tomography%20scanner&rft.jtitle=Oral%20surgery,%20oral%20medicine,%20oral%20pathology%20and%20oral%20radiology&rft.au=Koivisto,%20J.,%20PhLic&rft.date=2014-09-01&rft.volume=118&rft.issue=3&rft.spage=355&rft.epage=362&rft.pages=355-362&rft.issn=2212-4403&rft.eissn=2212-4411&rft_id=info:doi/10.1016/j.oooo.2014.05.016&rft_dat=%3Cproquest_cross%3E1556290011%3C/proquest_cross%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c411t-27487bedfec02d6e01c517d2830395c374d75fda8162764c279d4d771f92c87e3%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=1556290011&rft_id=info:pmid/25151590&rfr_iscdi=true