GeneSearch™ BLN Assay could replace frozen section and touch imprint cytology for intra-operative assessment of breast sentinel lymph nodes
Background Breast cancer sentinel lymph node (SLN) biopsy has become a common procedure. The GeneSearch™ Breast Lymph Node Assay is a real-time reverse-transcriptase polymerase chain reaction assay for detecting nodal metastases larger than 0.2 mm. The trial is a prospective multi-center clinical tr...
Saved in:
Published in: | Breast cancer (Tokyo, Japan) Japan), 2014-09, Vol.21 (5), p.583-589 |
---|---|
Main Authors: | , , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Background
Breast cancer sentinel lymph node (SLN) biopsy has become a common procedure. The GeneSearch™ Breast Lymph Node Assay is a real-time reverse-transcriptase polymerase chain reaction assay for detecting nodal metastases larger than 0.2 mm. The trial is a prospective multi-center clinical trial conducted to validate the assay in China.
Methods
Four hundred and seventy-nine consecutive prospective patients were enrolled from six centers. SLNs were sectioned along the short axis into multiple blocks. Odd blocks were tested by the assay intra-operatively, and even blocks were assessed by post-operative histology. Six 4- to 6-μm-thick sections were taken every 150 μm per block. In addition, intra-operative histological assessments were performed on the even blocks of 214 patients by frozen section (FS) and all blocks of 156 patients by touch imprint cytology (TIC).
Results
A total of 1046 SLNs were excised. Overall performance of the assay compared to post-operative histology was accuracy of 91.4 %, sensitivity of 87.5 %, and specificity of 92.9 %. There were no significant differences in assay performance of each center. After a learning curve of about 10 cases, the assay could be performed in a median time of about 35 min. The sensitivity of the assay was similar to the FS (84.9 %,
P
= 0.885) and was significantly higher than the TIC (70.0 %,
P
= 0.007) while the specificity of all were comparable.
Conclusion
The GeneSearch™ Breast Lymph Node Assay is an accurate and rapid intra-operative assay for breast SLNs and it can replace FS and TIC for application. |
---|---|