Loading…
Multi-site evaluation of the BD Stem Cell Enumeration Kit for CD34+ cell enumeration on the BD FACSCanto II and BD FACSCalibur flow cytometers
Abstract Background aims Evaluation of the BD Stem Cell Enumeration Kit was conducted at four clinical sites with flow cytometry CD34+ enumeration to assess agreement between two investigational methods: (i) the BD FACSCanto II and BD FACSCalibur systems and (ii) the predicate method (Beckman Coulte...
Saved in:
Published in: | Cytotherapy (Oxford, England) England), 2014-11, Vol.16 (11), p.1558-1574 |
---|---|
Main Authors: | , , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Abstract Background aims Evaluation of the BD Stem Cell Enumeration Kit was conducted at four clinical sites with flow cytometry CD34+ enumeration to assess agreement between two investigational methods: (i) the BD FACSCanto II and BD FACSCalibur systems and (ii) the predicate method (Beckman Coulter StemKit and StemTrol, Immunotech SAS, Beckman Coulter, Marseille Cedex 9, France). Methods Leftover and delinked specimens ( n = 1032) from clinical flow cytometry testing were analyzed on the BD FACSCanto II ( n = 918) and BD FACSCalibur ( n = 905) in normal and mobilized blood, frozen and thawed bone marrow and leucopheresis and cord blood anticoagulated with citrate phosphate dextrose, anticoagulant citrate dextrose—solution A, heparin and ethylenediaminetetraacetate, alone or in combination. Fresh leucopheresis analysis addressed site equivalency for sample preparation, testing and analysis. Results The mean relative bias showed agreement within predefined parameters for the BD FACSCanto II (−2.81 to 4.31 ±7.1) and BD FACSCalibur (−2.69 to 5.2 ±7.9). Results are reported as absolute and relative differences compared with the predicate for viable CD34+ , percentage of CD34+ in CD45+ and viable CD45+ populations (or gates). Bias analyses of the distribution of the predicate low, mid and high bin values were done using BD FACSCanto II optimal gating and BD FACSCalibur manual gating for viable CD34+ , percentage of CD34+ in CD45+ and viable CD45+ . Bias results from both investigational methods show agreement. Deming regression analyses showed a linear relationship with R2 > 0.92 for both investigational methods. Discussion In conclusion, the results from both investigational methods demonstrated agreement and equivalence with the predicate method for enumeration of absolute viable CD34+ , percentage of viable CD34+ in CD45+ and absolute viable CD45+ populations. |
---|---|
ISSN: | 1465-3249 1477-2566 1477-2566 |
DOI: | 10.1016/j.jcyt.2014.03.006 |