Determining the Instantaneous Effective Energy of x-Ray Tube Bremsstrahlung
A method is described for determining the instantaneous effective energy of x-ray tube brehmsstrahlung by means of two semiconductor detectors employing epitaxial GaAs structures and a measurement circuit, which together determine the effective energy with an error of 5% in the range 20-80 keV in th...
Saved in:
Published in: | Measurement techniques 2003-08, Vol.46 (8), p.806-809 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A method is described for determining the instantaneous effective energy of x-ray tube brehmsstrahlung by means of two semiconductor detectors employing epitaxial GaAs structures and a measurement circuit, which together determine the effective energy with an error of 5% in the range 20-80 keV in the presence of nonlinearity in the detector response.[PUBLICATION ABSTRACT] |
---|---|