Loading…

Determining the Instantaneous Effective Energy of x-Ray Tube Bremsstrahlung

A method is described for determining the instantaneous effective energy of x-ray tube brehmsstrahlung by means of two semiconductor detectors employing epitaxial GaAs structures and a measurement circuit, which together determine the effective energy with an error of 5% in the range 20-80 keV in th...

Full description

Saved in:
Bibliographic Details
Published in:Measurement techniques 2003-08, Vol.46 (8), p.806-809
Main Authors: Dvoryankin, V F, Dikaev, Yu M, Kudryashov, A A, Sokolovskii, A A
Format: Article
Language:English
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A method is described for determining the instantaneous effective energy of x-ray tube brehmsstrahlung by means of two semiconductor detectors employing epitaxial GaAs structures and a measurement circuit, which together determine the effective energy with an error of 5% in the range 20-80 keV in the presence of nonlinearity in the detector response.[PUBLICATION ABSTRACT]
ISSN:0543-1972
1573-8906
DOI:10.1023/A:1026177800820