A new X-ray scintillator for digital radiography

The authors report a new scintillator based on a transparent ceramic of Lu/sub 2/O/sub 3/:Eu. The material has an extremely high density of 9.4 g/cm/sup 3/ and a light output comparable to CsI:Tl. Its narrow-band emission at 610 nm perfectly matches the spectral response of charge coupled devices (C...

Full description

Saved in:
Bibliographic Details
Published in:IEEE transactions on nuclear science 2003-06, Vol.50 (3), p.297-300
Main Authors: Nagarkar, V.V., Tipnis, S.V., Miller, S.R., Lempicki, A., Brecher, C., Szupryczynski, P., Lingertat, H.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The authors report a new scintillator based on a transparent ceramic of Lu/sub 2/O/sub 3/:Eu. The material has an extremely high density of 9.4 g/cm/sup 3/ and a light output comparable to CsI:Tl. Its narrow-band emission at 610 nm perfectly matches the spectral response of charge coupled devices (CCDs). To enhance the spatial and contrast resolution, the authors have developed a special process to pixelate the scintillator and prevent the spread of light within the scintillator volume. The imaging performance of this pixelated device was evaluated using a thermoelectrically cooled CCD camera. The new scintillator is expected to play a major role in digital radiographic systems when readout technologies capable of taking advantage of the transparency are developed further.