Long-Term Outcomes of a New Model of Anatomical Hydroxyapatite-Coated Hip Prosthesis

Abstract This prospective study was designed to evaluate 196 Anatomique Benoist Giraud (ABG II) total hip arthroplasties which were implanted between September 1999 and December 2000. A minimum 11 years follow up was completed in 183 cases. The bearing surfaces were polyethylene–zirconia in 84 cases...

Full description

Saved in:
Bibliographic Details
Published in:The Journal of arthroplasty 2013-08, Vol.28 (7), p.1160-1166
Main Authors: Herrera, Antonio, MD, PhD, Mateo, Jesús, MD, PhD, Lobo-Escolar, Antonio, MD, PhD, Panisello, Juan José, MD, PhD, Ibarz, Elena, PhD, Gracia, Luis, PhD
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Abstract This prospective study was designed to evaluate 196 Anatomique Benoist Giraud (ABG II) total hip arthroplasties which were implanted between September 1999 and December 2000. A minimum 11 years follow up was completed in 183 cases. The bearing surfaces were polyethylene–zirconia in 84 cases, polyethylene–metal in 42 and ceramic–ceramic in 57. Changes in the femoral stem design, in relation to the previous ABG I model, have led to a significant improvement in stress-shielding. Polyethylene wear rate was lower by more than 50% compared with non-crosslinked polyethylene. Excellent and good results were obtained in 90.32% of cases, and implant survival was 98.39% at the end of follow-up.
ISSN:0883-5403
1532-8406
DOI:10.1016/j.arth.2012.06.033