Loading…
Reliability of manikin‐based studies: an evaluation of manikin characteristics and their impact on measurements of ventilatory variables
Summary Findings from manikin‐based studies on ventilation are commonly directly extrapolated to clinical practice. The aim of this study was to determine how the use of manikins affects measurements of ventilatory variables. We connected a lung simulator to a manikin, which was then ventilated at d...
Saved in:
Published in: | Anaesthesia 2015-08, Vol.70 (8), p.915-921 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Summary
Findings from manikin‐based studies on ventilation are commonly directly extrapolated to clinical practice. The aim of this study was to determine how the use of manikins affects measurements of ventilatory variables. We connected a lung simulator to a manikin, which was then ventilated at different inspiratory flows. We defined three experimental models to compare measurements of ventilatory variables between the mechanical ventilator and the lung simulator. Even when no leakage occurred, significant tidal volume deviations were observed; from a mean (SD) of 21 (2) ml to 49 (9) ml, and from 40 (4) ml to 88 (5) ml for invasive and non‐invasive ventilation, respectively (p |
---|---|
ISSN: | 0003-2409 1365-2044 |
DOI: | 10.1111/anae.13099 |