Flow injection spectrophotometric method for the speciation of aluminium in river and tap waters
An automatic flow injection method for aluminium speciation in water at the nanogram per millilitre levels of reported, based on a Pyrocatechol Violet chelation-ion-exchange method. Data are presented on total reactive aluminium, total monomeric aluminium and non-labile monomeric aluminium. The prop...
Saved in:
Published in: | Analyst (London) 1993, Vol.118 (9), p.1199-1203 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | An automatic flow injection method for aluminium speciation in water at the nanogram per millilitre levels of reported, based on a Pyrocatechol Violet chelation-ion-exchange method. Data are presented on total reactive aluminium, total monomeric aluminium and non-labile monomeric aluminium. The proposed method is sensitive (detection limit, 5 mu g/l), rapid and reproducible. The results obtained in the analysis of six river and tap water samples were consistent with those provided by the manual counterpart of the method. |
---|---|