A rapid method for measuring intraocular lens power in vitro with a focimeter

In this paper we describe a new method for measuring the intraocular lens (IOL) power using a focimeter, a negative ophthalmic lens and a saline solution (0.9% NaCl). To test this we measured the power of 58 different IOLs and we compared them with the power stated by the manufacturer. Despite the l...

Full description

Saved in:
Bibliographic Details
Published in:Experimental eye research 2015-11, Vol.140, p.190-192
Main Authors: García-Domene, Mari Carmen, Díez-Ajenjo, María Amparo, Peris-Martínez, Cristina, Navea, Amparo, Artigas, José María
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:In this paper we describe a new method for measuring the intraocular lens (IOL) power using a focimeter, a negative ophthalmic lens and a saline solution (0.9% NaCl). To test this we measured the power of 58 different IOLs and we compared them with the power stated by the manufacturer. Despite the limitations, the results show a good correlation. •We present a fast method for measuring intraocular lens power with a focimeter.•We avoided some problems described by the ISO (expensive, time-consuming or different depending on the lens design).•This method can improve the speed in future research and it could be used by groups on a low budget.
ISSN:0014-4835
1096-0007